Integration of Certified Reference Material (CRM)
Henrik Zetterberg University of Gothenburg Ingrid Zegers Joint Research Centre, European Commission
2
Classical traceability chain
Definition of the SI unit e.g. AAA, refractive index “Pure” protein Reference procedure, immunoassay ring trial CRM matrix material Value transfer 1: LC-MS, immunoassay Manufacturer’s working calibrator Value transfer 2: immunoassay Manufacturer’s product calibrator Immunoassay CRM material QC Routine sample 2
3
Traceability Chain Two MS methods formally certified
by JCTLM: -C12RMP1 -C11RMP9
4
Specific for Abeta42 Decision to produce 3 levels
Dilutional linearity not a given Handling issues (sticking to vials, tips, etc.) In-house and kit calibrators may not be commutable Decision to produce 3 levels
5
Use of the CRMs for Abeta42
Three CRMs released on 1st December 2017
6
CRM characterisation ERM-DA480/IFCC ERM-DA481/IFCC ERM-DA482/IFCC 0.45
Certified value [µg/L] Uncertainty [µg/L] ERM-DA480/IFCC ERM-DA481/IFCC ERM-DA482/IFCC 0.45 0.72 1.22 0.07 0.11 0.18
7
How should the CRMs be implemented?