Equipment Operation and Maintenance
2
Grid-controlled tube Designed to turn on and off rapidly, providing short precise exposures Use in fluoro and in capacitor discharged (condenser-discharge) mobile units Uses a molybdenum focusing cup
3
FLUOROSCOPY INPUT PHOSPHOR PHOTOCATHODE OUTPUT PHOSPOR LARGE DIAMETER
CESIUM IODIDE X-RAYS LIGHT PHOTOCATHODE LIGHT ELECTRONS OUTPUT PHOSPOR SMALLER DIAMETER ELECTRONS LIGHT
4
FLUORO CONT. FOCUSSING LENSES VIDICOM/PLUMBICON BEAM SPLITTING DEVICE
mA ABS AGC or ABC
5
# OF INPUT LIGHT PHOTONS
FLUX GAIN=# OF OUTPUT LIGHT PHOTONS # OF INPUT LIGHT PHOTONS MINIFICATION GAIN = Diameter of input (squared) diameter of output (squared BRIGHTNESS GAIN: minification x flux
6
Trifield Tubes 25/17/12 smaller diameter of input phosphor reduces: field of viewmagnification minification gain dimmer image
7
Larger film format greater the # of films =higher patient exposure
Cinefluoroscopy rapidly obtained photospot image (35mm film)15,30,60,90 frames per sec. cardiac cath Spot film 12 frames per sec smaller film size cassette loaded
8
FLUORO QUESTIONS #9,24,30,36,38,40,44,47,54 63, 69,70
9
AEC
10
Collimation and AEC Photodiode ion chamber light sensitive
under table, behind cassette ion chamber charge sensitive above bucky and cassette #10,14,25,27,34,81,92,102,113,
12
CIRCUITS
13
Self induction = autotransformer and choke coil
Mutual induction =High voltage transformer & filament circuit V = VS = NS VP= NP I = NS = IP NP=IS
14
Rectifiers Filament circuit rheostat filament circuit
15
Star and wye configurations
Motor and generators Rheostat Star and wye configurations transformer windings in 3-phase circuits 5,20,23,26,35,45,46 ,51,52,53,74,77
16
THREE PHASE/SINGLE PHASE
HU single phase=ma x time x kvp three phase/ 6 pulse ma x time x kvp x 1.35 three phase/12 pulse ma x time x kvp x 1.41
17
VOLTAGE RIPPLE 100% 13% 4 or 3% ?’s 35,64,96,97,99,105,106
Single phase three phase 6 pulse three phase pulse 100% 13% 4 or 3% ?’s 35,64,96,97,99,105,106
18
Tomography Thinner the cut the _________the angle 43
19
QUALITY CONTROL TESTS SPINNING TOP TIMER WIRE MESH SCREEN/FILM CONTACT
PENETROMETER KVP AND CONTRAST STAR PATTERN FOCAL SPOT SIZE CUTIE PIE ROENTGEN DOSIMETER WISCONSIN TEST TOOL KVP IN BEAM GEIGER RADIOACTIVITY
20
spinning top tests Spinning top tests
half wave/single phase = exposure time x 60 Full wave/single phase = exposure time x 120 Three phase –exposure time =arc size/360 timers and rectifiers dots and dashes(12 dots of 1/10sec, 24 dots of 1/5 sec, 6 dots for 1/20 sec) synchronous spinning top solid arc #18,28,41,50,57,61,66,65,68,
21
TUBE LIFE/HEAT LOAD #6,12,16,33,44,73,77,101,
22
Mobile radiography, CT, Misc
Capacitors- temporarily stores energy-no battery 2,3,4,22,37,42,55,58,71,78,87,88,120
23
DIGITAL Corectec
24
Misc. Light absorbing dye in screens reduce diffusion of fluorescent light