1
How to align lightlever optics
Mirror bracket Clamp screw Access hole LightLever Pickoff tube mounted above prism box
2
Pickoff tube pieces pickoff tube lens lens slider wrench
mirror bracket
3
To inspect optics, first loosen clamp screw
with 3/32” hex key
4
Remove pickoff tube Be careful not to touch pellicle against edge of hole
5
Inspect pellicle Film should be flat and smooth
Remove dirt with puff of air Do not touch film with fingers or brush
6
Remove mirror bracket with 0.035” hex key (0.9 mm)
7
Use wrench to remove slider
Twist and pull
8
Inspect lens for dirt
9
Reassemble pickoff tube
Be careful tube is not twisted in mount
10
Place pickoff back into hole
Be careful not to touch pellicle against edge of hole
11
Align mirror bracket straight up/down and tighten clamp screw
12
Remove mirror bracket
13
so bare tube is sticking out
14
Turn on trap laser and adjust control rods so that trap focus is in middle of TV screen
15
Place IR indicator card 9” away from end of tube
Wear laser goggles to protect your eyes
16
Beam should focus to a point 9”distant and 5” above the table
Fake spot shown Real spot not so bright
17
Move lens slider in/out with wrench to adjust 9” focal distance
18
Loosen clamp screw and pull pickoff tube in/out to adjust 5” focal height
but do not rotate tube which changes twist of pellicle
19
Beam spot moves Up/Down when pickoff tube slides In/Out
Make spot 5” above table slide tube in/out
20
Tighten clamp screw and replace mirror bracket
21
Place IR detector card inside instrument head
Spot should be visible near position detector Fake spot shown here Real spot not so bright
22
If spot is not visible try twisting position of mirror bracket until you see spot on card
23
Adjust LightLever mirror screws with 5/64” (2mm) hex wrench
24
so as to put beam inside PSD target
25
Move beam up/down and right/left using control rods
26
Problem : Strange triangle pattern in LL window
Problem : Strange triangle pattern in LL window. Solution: Be sure spot passes through middle of TV screen and adjust light-lever mirror screws to center spot on PSD target
27
Now spot is centered on detector so pattern moves up/down and right/left with control rods
28
If pattern is tilted more than 10 degrees try twisting pellicle relative to mirror
29
The End