1 MR Airway Pressure Device
Group Members Laura Sheehan Kevin Johnson Jon Cappel Noelle Simatic Client Dr. Victor Haughton Advisor Prof. Mitch Tyler

2 Chiari I Malformations

3 MR Background Tuner 90.7 104.1 B

4 Device

5 Normal Breathing (Holes Open) Valsalva (Holes Closed)
Device Normal Breathing (Holes Open) Valsalva (Holes Closed) Vacuum Compressed Air

6 Validation Device Precision MR Compatibility Protocol Development
Does the MR environment effect precision? Does scanning effect precision? MR Compatibility Does our device cause artifacts? Does our device degrade image quality? Protocol Development What’s the best way to perform this? Human/Animal Subjects Testing Final evaluation!!

7 Device Precision: Test
Computer Controlled Flow Pump Transducer DAQ Trigger Measured ~240 Periods of Square flow waveform Outside room  With Setup in Side Room Inside Bore  Normal Setup During Scan Normal Setup during a scan

8 Device Precision: Results

9 Device Precision: Results
SNR Measures Outside room: 40.99 Bore: 28.3 During Scan: 26.8 All less than 5% error Correlation Analysis Plotted Values against each other Pearson Coef (R=0.998) Slope of 1.01

10 MR Compatibility: Test
Image 1 Image 2 Noise With Device Without Device Artifact

11 MR Compatibility: Preliminary Results
Baseline With Device No obvious artifact but This noise level is unacceptable !!!!!

12 MR Compatibility: Future Work
RF Z104 Top 100 Talk Radio DAQ Device NPR Shielding (copper) Control Room MR Suite

13 Future Work MR Compatibility Anesthetized Patient Prototype
Additional Shielding Anesthetized Patient Prototype Protocol Development Human/Animal Subjects Testing IRB Approval Outreach Presentation Final Deliverables

14 as Chiari Patient “Little Suzie”
Questions? Laura as Airway Team Member Noelle as Chiari Patient “Little Suzie” Kevin as Mitch Tyler Jon as Vic Haughton, M.D. Kevin as MR Tech