ModWorX Pro – Calibration, Verification Upgrade
Version 4.2.0 2017 / 2018
2
Upgrades Calibration Reports now shows the calibration of the input in As Found/As-Left values. Current and previous verifications are stored for reporting purposes. This allows a Verification, Calibration, Verification cycle to be reported in a single download.
3
Verification The verification procedure has not changed, minimizing training needs on the new software. 2 Verifications are saved in the device (Current and Previous) Note that the user only can view the “Error (% of FS)” column while calibrating. If the user wishes to see that in a report (and the new mode is being used) it can be seen by using Scanner Data Manager software. ModWorX does not support this reporting feature.
4
Calibration The calibration procedure has not changed, minimizing training needs on the new software. The columns are labeled “Applied/As Left” and “As-Found”. Applied/As-Left is the measurement value applied to the sensor As-Left is value of the measurement based on the previous calibration.
5
Verify, Calibrate, Verify
Verification results after completing the Verify, Calibrate, Verify cycle.
6
Configuration Page No changes have been made to the configuration page for the inputs. The configuration pages show the current calibration and verification
7
Reporting Upgrades The download reports presents the current and previous verification data plus the current calibration values. Calibration and verification notes are also supported.
8
Program Options Users can select the calibration format
Applied/Measured Legacy approach As-Left/As-Found New approach, the default and recommended selection