The First Objective Audiometer (Wedenberg)
Cow Bell The First Objective Audiometer (Wedenberg)
2
Subjective Component The patient
3
Subjective Component The Operator
4
Sound STR Tympanometry OAE ECoG 5ms BERA 10ms CERA 500ms CNV ERA S U B
J E C T I V E A U D I O M E T R Y Sound STR OUTER & MIDDLE EAR INNER EAR BRAIN STEM CORTEX CEREB- RUM Tympanometry OAE ECoG 5ms BERA 10ms CERA 500ms CNV ERA
5
Hang STR Tympanometria OAE ECoG BERA CERA CNV ERA S Z U B J E K T Í V
A U D I O M E T R I A Hang STR KÜLSŐ és KÖZÉP FÜL BELSŐ FÜL AGY- TÖRZS CORTEX CEREB- RUM Tympanometria OAE ECoG BERA CERA CNV ERA
6
Hang STR Tympanometria OAE ECoG BERA CERA CNV ERA S Z U B J E K T Í V
A U D I O M E T R I A Hang STR KÜLSŐ és KÖZÉP FÜL BELSŐ FÜL AGY- TÖRZS CORTEX CEREB- RUM Tympanometria OAE ECoG BERA CERA CNV ERA
7
Hang STR Tympanometria OAE ECoG BERA CERA CNV ERA S Z U B J E K T Í V
A U D I O M E T R I A Hang STR KÜLSŐ és KÖZÉP FÜL BELSŐ FÜL AGY- TÖRZS CORTEX CEREB- RUM Tympanometria OAE ECoG BERA CERA CNV ERA
8
Hang STR Tympanometria OAE ECoG BERA CERA CNV ERA S Z U B J E K T Í V
A U D I O M E T R I A Hang STR KÜLSŐ és KÖZÉP FÜL BELSŐ FÜL AGY- TÖRZS CORTEX CEREB- RUM Tympanometria OAE ECoG BERA CERA CNV ERA
9
Hang STR Tympanometria OAE ECoG BERA CERA CNV ERA S Z U B J E K T Í V
A U D I O M E T R I A Hang STR KÜLSŐ és KÖZÉP FÜL BELSŐ FÜL AGY- TÖRZS CORTEX CEREB- RUM Tympanometria OAE ECoG BERA CERA CNV ERA
10
Objectiv audiometry Evoked Responses Reflexes STR Acoustic response OAE Electric Response Audiometry: ERA ECoG ms BERA ms MLR ms Steadystate response CERA ms CNV ca. 2 sec
11
Age-dependence in examinations
Type of examination ERA, OAE Stapedial reflex Reflexaudiometry Behaviour-audiometry Play-audiometry Pure-tone audiometry Speech-audiometry Supratreshold-audiometry
12
Tympanometry, STR, MFT Tympanograms A-type B-type C-type Loudspeaker
Tone-generator A-type Airpump Equalizer Ampli-fier Voltmeter Earcanal-Probe Amplifier Microphon Voltage Controller Tympanometry, STR, MFT B-type C-type Normal Reflexdecay
13
Loudspeaker Amplifier Tone-generator Air-pump Amplifier Equaliser Voltmeter Ear-canal-Probe Microphone Voltage Controller
14
Pressure [dPa] Pressure [dPa] Pressure [dPa] Type A Type B Type C
15
Normal STR-decay Pathological STR-decay
16
Hearing Pathway
17
Photosummation
18
Digitalis purpurea Finger Hut Fox glove Gyüszűvirág
19
A/D Conversion AVERAGING AVERAGING
2048 sample Channels Analog signal AVERAGING 2048 sample averaging Digital data
20
Minták száma Sampling Rate Mintavétel sűrűsége Number of Sampling
sample Sampling Rate 85 points 51 points 256 points Mintavétel sűrűsége Number of Sampling
21
Sound stimuli Sound stimuli Click01 click10 click23
One-cycle Sinus Halve-cycle Sinus rectangular biphase Click01 click10 click23 Sound stimuli Sound stimuli STOP Sinus tone without modulatio risetime = 0 Wavolvaso Rising Plateau decreasing T.burst t.burst4 -> longer Ms/division
22
Built in the computer Sound generator Ampli- fier Monitor/ Printer
AD-Conv Averager Ampli- fier Pre Amplifier
24
ECoG
26
Tchaikovsky B minor piano concerto Panpipe
27
Amplifier Loudspeaker Tape recorder
28
0.day Neomycin 4.day Neomycin 9.day Hearing aid
31
SP less than half as AP is normal
SP more than half as AP is recruitment
32
Advantages of ECoG: Exact information about the periphery No negative interaction of the general anesthesia Disadvantages of ECoG: Information is only about the periphery Invasive
33
BERA (ABR)
34
Intenzity-latency gram
BERA-thresholds
35
Retrocochlear lesion I –V IPL 4.2 +/- 0.3 <4.5 normal
>4.5<4.65 suspect >4.6 RL Retrocochlear lesion
37
Waves are successive missing
Retrocochlear lesion
38
Recruitment
39
Advantages of BERA : Exact information from the periphery No negative interaction of the general anesthesia Noninvasive Brainstem topo-diagnostic Precise objective hearing threshold (2-4 kHz) Drawbacks of BERA: Information is only from the periphery There is no frequency-specificity
40
BERA
42
MLR-Threshold MLR FFR BERA
43
MLR BERA
44
40 Hz Response
45
Advantages of MLR : Noninvasive Frequency -specificity Drawbacks of MLR: Slight interaction of general anesthesia Threshold is not precise Time consuming
46
CERA 500 ms
47
Advantages of CERA: Noninvasive Frequency -specificity Drawbacks of CERA: Strong interaction of general anesthesia Threshold is not precise Time consuming
48
CNV MMN
49
ON-Responses all together
50
BeforeTreat-ment After Treat-ment
SUDDEN DEAFNESS BeforeTreat-ment After Treat-ment AUDIOGRAM BERA DPOAE TEOAE
51
OAE + BERA TUMOUR
52
STEADY-STATE POTENCIALS AMFR
MASTER John&Picton1998 Steady-St. Resp
55
David Kemp 1978
56
Thank You Very Much for Your Attention!