Technetium-99 in Urine Tabitha Liebrecht & Buzz Vickery
Severn Trent Laboratories Richland 2800 George Washington Way Richland, Wa (509)
2
Advantages of Eichrom Method (Summary)
Time saving Waste reduction Yield
3
Method Summary (Step One - Urine Prep)
Sample Pour-up Prepare Sample
4
Method Summary (Step Two - Separations and Counting)
Acid Digestion/Transfer Prepare sample for TEVA disk Filter/Separation via TEVA disc Prepare sample for LSC
5
Ion Exchange vs. TEVA Disc
Ion Exchange method spike yield (1999) 300ml sample size w/ 85.04% yield TEVA disc method spike yield (2000) 300ml sample size w/ 95.23% yield Yield increase Average of 10% increase in spike recovery
6
Ion Exchange vs. TEVA Disc (continued)
tSIE 100% Lum 66.0% New method MDA 15% Contractual MDA = 35 pCi/L
7
Ion Exchange vs. TEVA Disc (continued)
Analysis time saving 48 hours vs 12 hours Cost Savings (11 samples/batch) 75% savings per batch (FTE)
8
Waste Summary 800ml vs 350ml of acidic waste
60% decrease in waste produced Waste disposal savings of $0.66 per sample analyzed
9
Conclusions Improved TAT Decreased waste Increased yields Lower MDA
Improved reliability
10
Future Eichrom Projects
Np-237 by TEVA with ICP/MS (via Eichrom ACW08 and ACU03) Th separation Am/Pu/U by TRU/UTEVA (via Eichrom ACW03) Strontium by Sr resin (via SRW01) Tc-99 in soil Pb-210 Single Operator Test