Download presentation
Presentation is loading. Please wait.
1
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Gas Mixing Device
2
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Common Configuration LAB-manager2 basis (6 HE table box) © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps Lower casing: LAB-manager control unit
3
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Common Configuration LAB-manager2 basis (6 HE table box) © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps Upper casing: Mass flow controllers, gas attachments and connections
4
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Operation conditions Inputs: 1 to 4 gas inputs Outputs: 1 to 4 gas outputs Connection with Swagelok Quick couplers DN 4 Gas settings ex factory: Air, O2, N2, CO2 (not condensing) further gase on request Pressures: Input pressure 2 to 8 bar Output pressure 0 to 6 bar Pressure difference: min. 2 bar © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps
5
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Mass Flow Controller Thermal mass flow controller for air, O2, N2: 0 - 0,05 sLm, 0 - 0,5 sLm, 0 - 5 sLm for CO2: ,03 sLm, 0 - 0,3 sLm, 0 - 3 sLm sLm = standard liter / min (0°C, 1013 mbar) Each component is controlled seperately on each output © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps No multiplex operation!
6
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Control Unit analog and digital in- and outputs interface cards © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps power supply V, Hz, 2 A LAB-manager
7
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Gas mixing Unit Pressure transmitters © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps Mass flow controller Magnetic valves
8
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Other versions for stand-alone operation Stand-alone operation possible (based on LAB-Vision) GMix41 GMix44 © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps
9
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
User interface in RAMOS integrated fully integrated in RAMOS user interface © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps
10
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
User interface in RAMOS integrated GMix software runs on RAMOS desktop PC parallel in the background © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps
11
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Comfortable Handling Settings in RAMOS software operating GMix-software runs invisible in the background Operation- and error messages are logged in RAMOS documentation Monitoring of limit values Shut-off of the gas mixing device at input pressure > 8 bar, output pressure > customer limit value Warning at pressure differences < 2 bar (input/output) (required for safe flow control) Alarm messages at flow beyond tolerance © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps Inputs are pressure-resistant up to 20 bar
12
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Examples Example 1: RAMOS-high flow at 800 ml/min (standard) Mixture of gases: Luft: 95% (0,5 sLm + 0,5 sLm) CO2: 5% (0,5 sLm + 0,05 sLm) Controller setting: Air: ,00 ml/min Setting (0,5 sLm): 76 % of setting range ,00 ml/min Setting (0,5 sLm): 76 % CO2: ,00 ml/min Setting (0,5 sLm): 13 % © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps
13
© HiTec Zang GmbH - HRE - 25/07/2005 - GMix41.pps
Examples Example 2: RAMOS-high flow reduced to 600 ml/min, High flushing phase (between 40 and 80 s) extended accordingly Mixture of gases (synthetic): N2: 75% (0,5 sLm) O2: 20% (0,5 sLm) CO2: 5% (0,5 sLm + 0,05 sLm) N2: ,00 ml/min Setting (0,5 sLm): 90 % of setting range O2: ,00 ml/min Setting (0,5 sLm): 24 % CO2: ,00 ml/min Setting (0,05 sLm): % © HiTec Zang GmbH - HRE - 25/07/ GMix41.pps