Attaching the Micromanipulator to the Laser
Laser In-Service Attaching the Micromanipulator to the Laser
2
Micromanipulator Suitcase
2 Suitcases Found on Laser Cart “Coherent” “Lumenis”
3
There are 2 colposcopes – 1 “old” and 1 “new” – the Micromanipulator connects differently to each one. Usually at WBG Bunting Old Colposcope New Colposcope **Only the old colposcope has a connection for a camera head**
4
Step 1 Remove Objective Lens from Colposcope Old Colposcope
New Colposcope Remove Objective Lens from Colposcope Leave Objective Lens in Place on colposcope--- Guide Silver Plate into grooved square on colposcope & tighten screw
5
Step 2 Attach black ring labeled “Part 1” into colposcope
Old Colposcope New Colposcope Attach black ring labeled “Part 1” into colposcope Attach black ring labeled “Part 1” into silver plate
6
Step 3 Attach 2nd black ring (Clearspot) into “Part 1” ring
Old Colposcope New Colposcope Attach 2nd black ring (Clearspot) into “Part 1” ring Re-Attach Objective lens (300)
7
Step 4 Attach 2nd black ring (Clearspot) Old Colposcope New Colposcope
Attach Micromanipulator – by loosening screw and snapping it onto the 2nd black (Clearspot) ring, then re-tighten screw. Ensure it is secure.
8
Step 5 Attach Laser to black end of Micromanipulator Old Colposcope
New Colposcope Attach Micromanipulator – by loosening screw and snapping it onto the 2nd black (Clearspot) ring, then re-tighten screw. Ensure it is secure. Attach Laser to black end of Micromanipulator
9
Step 6 Old Colposcope Attach Laser to black end of Micromanipulator
10
Focusing 1. Position the Focusing Lever to Index 1
11
2. Position a focusing target in front of the Micromanipulator at the appropriate working distance.
3. Adjust the distance between the eyepieces so that a single image is viewed in the field Adjust eyepiece here.
12
4. Set the image magnification to the lowest setting and grossly focus on the desired target.
13
5. Set the image magnification to the highest setting and fine focus on the desired target.
14
Reminders -Make sure objective lens and focal distance selector are on the same power setting (300) Test the laser Drape with wet towels
15
-FILL OUT THE LASER LOG -RETURN THE LASER KEY TO THE PROPER LOCATION
-MAKE SURE THE LASER ARM IS IN THE “UP” POSITION -FILL OUT THE LASER LOG -RETURN THE LASER KEY TO THE PROPER LOCATION