MEG実験の液体Xe検出器について 東大 ICEPP 森研究室 M1 金子大輔
2
What is needed for MEG experiment ?
→ Reduce backgrounds as much as possible → High accuracy g-ray detector ! ・ Energy 53 MeV ・ Timing 0.1 ns ・ Position 4 mm We adopted Liquid Xenon detector
3
Merits and Difficulty about XENON
GOOD BAD Xenon emits much scintillation light (about 75% of NaI(Tl)) Decay time is very short ~ 45 ns Radiation length is short ~ 2.77 cm ↑(Z=54) Uniformity able to purify Xenon is liquid only in K ~ 165K Latent heat is small Insulation vessel is needed
4
Light emission process of xenon
There are two ways of scintillation process lpeak= 178nm (vacuum ultra violet ray) no self absorption ! Xe e- ionization Xe Xe+ Xe2+ e- excitation Xe* Xe Xe Xe** Xe t = 45 ns Xe* recombination Xe2* 1S+u 3S+u t = 4.2ns t = 22ns Xe Xe + photon
5
Our liquid xenon detector
>850 litter volume liquid xenon “C” shape to fit magnet 846 photo-multiplier-tubes (PMT) observe scintillation light Pulse Tube Refrigerator control xenon’s temperature To reduce mass of inner face honeycomb window is equipped
6
Connection around xenon detector
storage tanks gas purifier 1000L tank liquid purifier
7
g Principles of detection and analysis
PMTs catch scintillation photons and output signals Each quantity is calculated Energy of g ← sum of signals of all PMTs Time ← mean of timings Position ← light distribution g wave form from a PMT
8
Detector’s Outside high-voltage connector signal connector
COBRA-magnet honeycomb window
9
Detector’s Inside PMTs work in Low temperature (~160K)
magnetic field (<50 Gauss) LEDs for callibration are attached wires & plates with radio-active source (Am)
10
Photo Multiplier Tube (PMT)
Quantum efficiency (Q.E.) ~ 14 % High Voltage ~ 800 V , gain ~ 2×106 2 in. f HAMAMATSU R9288 features Metalchannel-dinode → short body → not to block gray from backward ・ tough to magnetic field quartz window →transparent to VUV Al strip on window → reduce resistance in low temperature Cathode = K-Sc-Sb
11
Pulse Tube Refrigerator
KEK Haruyama et.al stable temperature & pressure control within DT~0.1K , DP~0.001MPa How pulse tube refrigerator work ? Cooling power 200W 165K
12
Xenon Purification System
Water and oxygen mainly absorb scintillation light ↓ MEG detector have two ways of purification systems gas purification liquid purification absorb water more efficiently than gas absorb O2 H2O etc, purifier Finally labs → 300cm getter
13
Electronics , DAQ DRS on-line computers detector waveform digitizer
off-line computers splitter trigger on VME boards every event, waveforms of all PMTs are saved. Domino Ring Sampler (DRS) sample rate 2 GHz trigger board digitalize signals with FADC and FPGA slow control MSCB IN 2007 Runs 9974 Total TB
14
Calibration methods alpha source (241Am) → Q.E. , attenuation length
LED → gain , timing C-W proton beam → g energy Charge exchange → g energy etc ↑PMT gain calculation from LED runs ↑position detection from a
15
gray from nucleic or particle reaction is used for energy scale calibration
detector PSI cyclotron Proton beam Cockcroft-Walton (C-W) accelerator target muon beam pion beam Charge exchange reaction P + p- → n + p0 p0 → g + g , 83 MeV radiative capture P + p- → n + g MeV Li ( p , g ) Be 17.6 , 14.6 MeV B ( p , g ) C 4.4 , 11.7 , 16.1 MeV
16
Thanks for your listening !!
Status and Prospects 55 MeV g sup = 2.4% FWHM = 6.5% NOW analyzing data improving detector APRIL 2008 ~ Data taking again Thanks for your listening !!