Artifacts and Pitfalls in MR Angiography (MRA) for Pulmonary Embolism Meyer CA, Schiebler ML, Reeder SB, Francois CJ, Nagle SK
2
Disclosures Financial: UW has a research agreement with GE Healthcare
Only gadofosveset (Ablavar®) is FDA approved for MRA Use of other gadolinium based contrast agents is not FDA approved for MRA
3
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
4
MRA Technique Multiphasic acquisition Full chest coverage
fluoro triggered arterial phase immediate post-injection phase 2nd post-injection “steady state” phase using lower flip angle each phase done during single breath hold Full chest coverage High spatial resolution 2D parallel imaging must use multichannel phased array coil recommend ≥ 8 channels Freq encoding Phase encoding Slice encoding
5
MRA Technique k-space “corner cutting” Elliptical centric acquisition
Freq encoding Phase encoding Slice encoding k-space “corner cutting” eliminates 22% of readouts Elliptical centric acquisition enables fluoro-triggering 0.1 mmol/kg of gadobenate dimeglumine typically 15-20mL diluted with saline to a total volume of 30 mL
6
MRA Technique
7
Slab Orientation Frequency encode long axis of slab (S/I)
Sagittal slab excitation to minimize aliasing Freq encode arm torso
8
Pulmonary emboli Findings:
Filling defects in pulmonary arteries (arrows) Perfusion defects (arrowheads)
9
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
10
Bolus Timing ~15 sec elliptical centric acquisition
0.1 mmol/kg dose injected at 1.5 mL/s 200 lb patient: ~20 mL dose 13 s bolus 100 lb patient: ~10 mL dose 7 s bolus The Problem: bolus duration < acquisition time The Solution: extend bolus duration by diluting contrast in saline to 30mL injected at 1.5 mL/sec (20 s bolus) Is this bolus timing or bolus duration? Maki, et al., JMRI 6(4):642-51, 1996.
11
Bolus Timing Acquisition Enhancement Time inject
Contrast plateau leads to sharper vessels due to higher signal at edge of k-space Enhancement Dilute Bolus Standard Bolus Time inject
12
Scan late or bolus short
Scan early Ideal timing k0 kmax k0 kmax k0 kmax Artifacts will not persist on repeat acquisitions or repeat injection PA PA Blurred PA Aorta Aorta Aorta Edge enhanced
13
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
14
Contrast Injection Pitfalls
Transient Interruption of Bolus* Flow-related phenomenon Related to valsalva during breath-hold Increased unopacified blood return from the IVC Dual Injector Error Inadvertent reversal of the saline and contrast syringe Fluoro-triggering detects the small amount of contrast mistakenly used to “flush” the IV line during set up. Results in scanning when primarily saline is intravascular * Wittram C, Yoo AJ. J Thorac Imaging 2007; 22: 125-9
15
Dual Injector Error Small amount of contrast in aorta
Dense contrast in subclavian vein Small amount of contrast in aorta No contrast in pulmonary artery Pre-Injection During Injection 1st Post Injection 2nd Post Injection Need description
16
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
17
Gibbs Ringing Occurs near high-contrast edges
Occurs centrally in vessels 3-5 pixels wide based on true resolution May be mistaken for pulmonary embolism Measure signal drop within vessel If < 50% signal drop, suspect artifact If > 50% signal drop, suspect embolus Do not aggressively window vessels Ensure that background noise is visible in image
18
Gibb’s Ringing Pulmonary Embolism Simulated true vessel cross-section
Simulated Gibb’s ringing Actual cross section
19
Corner Cutting With Corner Cutting Without Corner Cutting
2 pixels 4.5 pixels 7 pixels Sagittal excitation SI frequency-encoding AP & RL phase-encoding With and without corner-cutting 1 x 1 x 1 mm3 resolution 2 NEX No parallel imaging With corner cutting, the point-spread function is circularly symmetric and Gibb’s ringing looks more like true embolism. Without corner cutting, there is less risk of misinterpreting Gibb’s ringing as true embolism.
20
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
21
Wrap and Parallel Imaging
Wrap (aliasing) occurs if excited tissue extends outside FOV sagittal slab excitation avoids wrap from arms Parallel imaging propagates wrap artifacts into the center of the image must completely include AP dimension of patient Don’t rely only on mid-sagittal and mid-axial scout images! Largest AP dimension usually at breasts or belly image noise worse in center of image (increased G-factor) Solution: Increase number of AP slices while maintaining reasonable breath-hold time, even at the cost of lower AP resolution
22
Residual Aliasing & G-factor
Wrap Parallel imaging Wrap
23
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
24
Amplifier Over-ranging
Problem: Prescan adjusts amplifier gain to use the entire dynamic range of the analog to digital converter Actual signal intensity of the acquisition may exceed this range due to IV contrast Solution: Decrease amplifier gain and reinject If this is a regular problem, then routinely decrease amplifier gain during manual prescan prior to injection. True k-space Over-range portion True image Over-range portion Observed Image
25
Amplifier Over-ranging
26
Outline MRA Technique MRI Reconstruction Patient Contrast Related
Bolus Timing Respiratory Motion Gibbs Ringing Bolus Duration Wrap and Parallel Imaging Transient Contrast Interruption Amplifier Over-ranging Dual Injection Error
27
Respiratory Motion Artifact
Patient’s often dyspneic if PE suspected If motion occurs in the middle of K space smearing occurs Injection startle motion quiescent Solution: Multiphasic injection – motion common on arterial phase (1st) acquisition Repeat injection with fewer, thicker slices to shorten acquisition time
28
Respiratory Motion Artifact
1st Injection has respiratory motion 2nd Injection
29
Conclusion Contrast-enhanced MRA is a mature and robust technology
High quality scans require careful attention to k-space sampling strategies, injection protocols, and technologist training Accurate interpretation requires under-standing common pitfalls and artifacts