1 MacDermid Plc Offshore Div.
Particle Analysis And Contamination Control Simon McManus

2 INTRODUCTION WHY PARTICLE COUNTING IS CARRIED OUT WHAT ARE PARTICLES?
AUTOMATIC PARTICLE COUNTERS CALIBRATING A MICROSCOPE COUNTING CONCLUSION

3 WHY PARTICLE COUNT? PREVENTATIVE MAINTENANCE EXAMINATION OF DEBRIS
CHANGES IN CONTAMINATION LEVEL PREDICT FILTER BLOCKAGE AND SYSTEM FAILURE (SEALS, PUMPS AND VALVES)

4 WHAT ARE PARTICLES? AN INDIVIDUAL UNIT OF PARTICULATE MATTER
SIZE AND SHAPE VARY GREATLY WHAT THEY ARE MADE FROM HOW THEY HAVE BEEN FORMED INTRAMOLECULAR FORCES PRESENT

5 DAMAGE EFFECT FROM PARTICLES
SOFT PARTICLES (PASTE, FILTER MEDIA, SEALS ETC.) CAN BLOCK FILTERS. HARD PARTICLES CAUSE ABRASION AND WEAR. THESE ALSO BLOCK FILTERS BUT TO A LESSER EXTENT. HARD PARTICLES INCREASE CONTAMINATION LEVELS.

6 Polymers and Fibres

7 Black Metal

8 Bright Metal

9 Rust

10 Silica

11 AUTOMATIC PARTICLE COUNTERS (APC’s)
LIGHT SCATTERING OR OBSTRUCTION USE LIGHT REACTIVE ELECTRONICS TO SIZE THE SHADOW OF PARTICLES OTHERS USE PRESSURE DIFFERENTIAL ACROSS MESH FILTERS OF A SET SIZE

12 PROBLEMS WITH APC’s IRREGULAR PARTICLES ARE ‘ROUNDED’ TO AVERAGE SIZE
LONG FIBRES ARE SEEN AS A STRING OF SMALL PARTICLES GAS BUBBLES AND INSOLUBLE LIQUIDS ARE SEEN AS PARTICLES APC’s HAVE PROBLEMS WITH OCEANIC HW 500 FLUIDS

13 CALIBRATING A MICROSCOPE
THE MICROSCOPE SHOULD HAVE A GRATICULE SCALE THE GRATICULE MUST BE CALIBRATED A CALIBRATED STAGE MICROMETER SHOULD BE USED IN REALITY THIS IS NOT ALWAYS POSSIBLE

14 FIELD CALIBRATION OF GRATICULES
TOTAL FILTER MEMBRANE AREA SHOULD BE KNOWN. STANDARD MEMBRANES ARE 960mm squ EACH INTERIOR SIDE OF ONE SQUARE IS 308µm.

15 PARTICLE ANALYSIS. ARP598B
100 ml OF FLUID IS USED A MICROSCOPE SLIDE IS PREPARED (FILTER MEMBRANE) AREAS OF THE SLIDE ARE EXAMINED AND CONTAMINATION LEVELS ARE MATHEMATICALLY CALCULATED FOR THE WHOLE SAMPLE SIZES OF 5, 15, 25, 50 AND OVER 100 µm ARE EXAMINED

16 GENERAL RULES DURING COUNTING
TRY TO EXAMINE AREAS OVER THE WHOLE MEMBRANE, DO NOT CONCENTRATE ON THE MIDDLE FEW SQUARES USE THE LARGEST DIMENSION OF THE PARTICLE FIBRES ARE SIZED USING THEIR LENGTH. A FIBRE HAS A LENGTH MORE THAN 10 X ITS WIDTH.

17 TAKING A SAMPLE TURN ON THE SAMPLE VALVE TO OBTAIN A STEADY FLOW
LET HALF A LITRE OR SO FLUSH THE VALVE FLUSH THE SAMPLE BOTTLE AND LID AT LEAST TWICE IN THE STREAM OF FLUID DO NOT TOUCH THE VALVE UNTIL THE SAMPLE HAS BEEN TAKEN.

18 SAMPLE PREPARATION SAMPLE HANDLING WILL ALWAYS INCREASE CONTAMINATION LEVELS SAMPLING WILL ALWAYS INCREASE CONTAMINATION LEVELS THE CONTAMINATION IN THE SYSTEM IS ALWAYS LESS THAN THE COUNT THE ERROR SHOULD BE KEPT TO A MINIMUM

19 SAMPLE CARE IF SAMPLES ARE TO BE PACKAGED CHOSE THE PACKING MATERIAL CAREFULLY AVOID BOTTLE LIDS WITH PAPER INSERTS CLING FILM OVER THE TOP OF THE BOTTLE CAN PREVENT DIRT INGRESS

20 CARE DURING PREPARATION
SLIDES SHOULD BE PRODUCED IN A CLEAN ROOM DO NOT BREATH ON OR LEAN OVER THE SAMPLES MAKE SURE ALL THE EQUIPMENT THAT THE FLUID MAKES CONTACT WITH HAS BEEN FLUSHED.

21 BLANK ANALYSIS PREPARE A BLANK SLIDE USING FILTERED WATER, 100ml +or- 5ml THIS COUNT WILL GIVE AN INDICATION OF BACKGROUND CONTAMINATION THE TOTAL COUNT SHOULD NOT EXCEED 10% OF THE LEVELS BEING INVESTIGATED.

22 COUNTING PROCEDURE IF DEBRIS DISTRIBUTION IS RANDOM, UNEVEN DISTRIBUTION WILL INVALIDATE THE COUNT. COUNTING IS DONE TO ARP 598 TO GIVE A NAS OR AS CLEANLINESS LEVEL. ERROR IN COUNTING SHOULD NOT BE GRATER THAN 10% BETWEEN OPERATORS OR LABORATORIES

23 NAS 1638

24 AS4059 Counts are reported as cumulative rather than interval
Optional Smaller size code Numeric Cleanliness Class in a specified Letter coded size.

25 AS4059