Presentation is loading. Please wait.

Presentation is loading. Please wait.

IgG Detection Using Anti-IgG Conjugated Microparticles

Similar presentations


Presentation on theme: "IgG Detection Using Anti-IgG Conjugated Microparticles"— Presentation transcript:

1 IgG Detection Using Anti-IgG Conjugated Microparticles
By Sarah Leung and Anbar Najam

2 Methods Conjugated anti-IgG antibodies to 0.92 um microparticles
33% microparticle surface coverage with IgG 0.02% microparticle concentration in solution

3 Methods Positioned a 180 degree optical reflection probe directly above samples for detection on a two-well slide

4 Results 5*10-11 g/mL to 5*10-6 g/mL IgG serial dilutions

5 Results 5*10-15 g/mL to 5*10-6 g/mL IgG serial dilutions

6 Discussion Results lacked evidence of any trends between antigen concentration and light scattering intensity Possible explanations: Coagulation of microparticles High scattering due to 180 degree detection Conjugation of antibody, rather than antigen, to microparticles

7 Next Experiment Added TWEEN 80 to existing 0.92 um microparticle solution and evaluate 180 degree vs. 45 degree light scattering detection Vs. 180 degrees 45 degrees

8 Results 5*10-14 g/mL to 5*10-6 g/mL IgG serial dilutions

9 Results 5*10-14 g/mL to 5*10-6 g/mL IgG serial dilutions

10 Discussion 180 degree results lacked evidence of any trends between antigen concentration and light scattering intensity With 45 degree probe, we found a detection limit of g/ml

11 Contact Angle Measurements of Anti-E. Coli and Anti-E
Contact Angle Measurements of Anti-E. Coli and Anti-E. Coli Conjugated Microparticles on Superhydrophobic Surface

12 Methods Conjugated anti-E. coli antibodies to 0.92 um microparticles
33% microparticle surface coverage with anti-E. coli 0.02% microparticle concentration in solution

13 Methods Dispensed 10 ul droplet on to superhydrophobic surface.
Measured contact angle over 10 minutes Droplets: Deionized Water Anti-E. coli -Conc. 5*10-7 Anti-E. coli conjugated to microparticles

14 Results Anti-E. coli -Conc. 5*10-7

15 Discussion Results oscillate and do not form a smooth curve.
Possible explanations: Surface contamination Droplet movement, vibration. Position of droplet. Size of droplet

16 Next Experiment Decrease the volume of the droplet from 10 ul to 3 ul for a more spherical shape.

17 Methods

18 Results Deionized Water

19 Future Work Minimize the oscillations.
Change the position of the surface. Determine an efficient way to dispense 3ul droplet on to superhydrophobic surface

20 Acknowledgement s UA/NASA Space Grant Dr. Jeong-Yeol Yoon
Lab group: Tremaine B. Powell, Jin-Hee Han, Brian C. Heinze, Phat L. Tran, Jennine N. Chesler, Dr. Keesung Kim, and Dr. Song


Download ppt "IgG Detection Using Anti-IgG Conjugated Microparticles"

Similar presentations


Ads by Google