Long-term results of balloon-expandable LifePath endografts in abdominal aortic aneurysm: A single-center experience Stefan Ockert, MD, Hendrik von Tengg-Kobligk, MD, Tanja Kippenhan, Patrick Kurz, MD, Dittmar Böckler, MD, PhD, Hans-Henning Eckstein, MD, PhD, Hardy Schumacher, MD, PhD Journal of Vascular Surgery Volume 50, Issue 3, Pages (September 2009) DOI: /j.jvs Copyright © 2009 Society for Vascular Surgery Terms and Conditions
2
Fig 1 Kaplan-Meier survival analysis for generation I (blue) and generation II (red) LifePath devices. For generation I, the standard error never exceeded 10%; for generation II, the standard error exceeded 10% after 24 months. Journal of Vascular Surgery , DOI: ( /j.jvs ) Copyright © 2009 Society for Vascular Surgery Terms and Conditions
3
Fig 2 Kaplan-Meier analysis for freedom of endoleak. The standard error never exceeded 10%. Journal of Vascular Surgery , DOI: ( /j.jvs ) Copyright © 2009 Society for Vascular Surgery Terms and Conditions
4
Fig 3 Top-wire form fractures (arrows).
Journal of Vascular Surgery , DOI: ( /j.jvs ) Copyright © 2009 Society for Vascular Surgery Terms and Conditions
5
Fig 4 Conventional radiographs show (a) overhead view and (b) zoomed view of wire form fractures (arrows). Journal of Vascular Surgery , DOI: ( /j.jvs ) Copyright © 2009 Society for Vascular Surgery Terms and Conditions