1 MuCool Test Facility Instrumentation Update
M. Haney, D. Errede, Z. Conway University of Illinois at Urbana-Champaign 12/13 August 2002

2 Instrumentation Seeking an intrinsically safe solution for instrumenting the LH2 absorber in the MTF Prudent to accommodate MICE and European safety considerations i.e. do it right the first time… 12/13 August 2002 m. haney, et al.

3 Objective - Safety Intrinsic safety, if possible
Limited power per channel Good idea for Fermi; required for Europe Gas-purged enclosure, if necessary 19” rack, 40U tall 12/13 August 2002 m. haney, et al.

4 Instrumentation (1) Lakeshore 218S TG-120PL GaAlAs diodes (8)
8 channel cryo temperature monitor +/- 21m K at 10 K 16 readings/s GPIB interface (to PC) TG-120PL GaAlAs diodes (8) -180mV/K at 4.2 K B < 5 T; rad hard 12/13 August 2002 m. haney, et al.

5 Instrumentation (2) FISO Fiber-optic strain and temperature
BUS chassis, with 4 (up to 8) modules 1000 readings/s RS-232 interface (to PC) FOS-N strain sensors +/- 500 me; 0.01% full scale; 0.2mm O.D. FOT-L temp sensors 0.1 K resolution; 1.5mm x 32mm (10mm active) Slow: 1.5 second response… 12/13 August 2002 m. haney, et al.

6 Instrumentation (3) Gateway E-4000 PC PCI-MIO-16E-1
1.8 GHz, 1Gbyte RAM, 0.5Gbyte cache 120Gbyte disk, Windows 2000 15” LCD flat panel display 640x480 CCD camera, microphone PCI-MIO-16E-1 16 channel ADC, 1.25 Msample/s, 12 bits, +/-10 V Tripplite Smart UPS 3000 VA 12/13 August 2002 m. haney, et al.

7 Instrumentation (4) IRM 64 channel multiplexed ADC
16 bit, 100 Ksamples/s ACNET network connection All PC data will be posted to IRM for ACNET access 12/13 August 2002 m. haney, et al.

8 Instrumentation (5) MTL7055ac barriers MTL7060ac barriers Low level AC
24 W per line 3 V max MTL7060ac barriers Star-connected AC 101 W per line 8.5 V max 12/13 August 2002 m. haney, et al.

9 (basic) Barrier fuse Power + Signal “safe circuit” Return Ground
12/13 August 2002 m. haney, et al.

10 Signals inside (the cryostat)
8 (cryo) temperatures readout via GPIB + LabView (PC) Multiple fibers FISO strain, temperature (4) Pressure, flow? Laser occlusion (2) Simple spark detectors (2) (light pipe to photodiode) 25(ish) 4-wire bolometry strips 12/13 August 2002 m. haney, et al.

11 (more) Signals inside Piezo vibration sensor
via (16 chan/12 bit) ADC and LabView (PC) Local (64 chan/16 bit) IRM channels via “LabView” (see last slide) MICE channels (?) placeholder for the future… 12/13 August 2002 m. haney, et al.

12 Signals outside USB (4 wires, and shield) for CCD camera and microphone via PC Many voltages, currents, temperatures, pressures magnets, beam position monitors, etc. Most available from ACNET or local IRM via “LabView” (see last slide) or Cyro-PLC 12/13 August 2002 m. haney, et al.

13 Two Possibilities (1) Intrinsically safe Hazard Safe Sealed Conduit(s)
Barrier(s) ACNET IRM Cryo(temp) FISO PC w/16 chan ADC network power Intrinsically safe signal conditioners and transmitters UPS 12/13 August 2002 m. haney, et al.

14 Two Possibilities (2) Gas-purged box Hazard Safe Barrier(s) Sealed
Conduit(s) ACNET IRM Cryo(temp) FISO PC w/16 chan ADC network power UPS 12/13 August 2002 m. haney, et al.

15 This Summer – Z. Conway LabView interface Integration – remains!
GPIB to cryo temp monitor - works PC-internal ADC (e.g. for piezo) – “works” Simple ADC data read - works High rate collection and compression – not yet IRM communications IRM-ADC Read - works Write – any moment now… Integration – remains! 12/13 August 2002 m. haney, et al.

16 Open Issues Wire+Shielding Concerns
Noise/sensitivity issues due to wire+barrier resistance? Approx 12 W /ft for 36 AWG manganin wire 48 W to 202 W added by barrier And diode capacitance… Common mode (surges) due to magnets? Need to protect electronics without burning barriers Cable plant into solenoid 12/13 August 2002 m. haney, et al.

17 Open Issues (2) Software Concerns Posting data to the IRM
For ACNET access Reading data from ACNET For PC access Remote control of PC Citrix ICA, PCAnyWhere, etc. LabView web-remote instrumentation KVM? 12/13 August 2002 m. haney, et al.

18 Open Issues (3) FISO strain and temp sensors
Not official rated below -40 C (!) Temp has been used at 35 K (custom) No experience with strain gages Company very willing to work with us To adjust signal conditioner, correction curves, etc. Will require some form of agreement Confidentiality, exclusivity for the instrument Yet freedom to use and publish measurements Also pressure, flow? 12/13 August 2002 m. haney, et al.