Artifact analysis
2
Artifact analysis One of the causes of rejected images listed in the repeat analysis worksheet is the presence of image artifacts. An artifact can contribute significantly to the total repeat rate; therefore a thorough knowledge of artifacts and their possible causes is necessary so that corrective action can be taken.
3
What Is An Artifact? Any optical density that is not anatomy
An unwanted density on the radiograph Interferes with diagnosis
4
There Are Three Time Periods In Which Artifacts Can Occur
During exposure During processing During film handling, prior to processing, loading cassette, or storing
5
Exposure Artifacts They are caused by the patient, the technologist, or the equipment during a diagnostic procedure Poor screen-film contact Improper use/positioning of a grid Double exposure Patient motion Poor patient preparation Clothing, jewelery, glasses, belts, etc.
6
Processing Artifacts They are caused by or occur during the processing of diagnostic images Dirty Rollers Will leave deposits on film Areas of increased or reduced density Occasionally some sludge may not be washed off and dried onto the film Chemical Fog Improper or inadequate chemistry Leaves a uniform dull gray haze on the film May also be seen in two different colors Referred to as dichroic stain
7
Guide Shoe and Roller Marks
Guide shoes are sprung or improperly positioned The ridges in the guide shoe press against the film leaving an artifact Pi-lines Dirt or stain on a roller 3.14 inches apart Pressure marks Irregular or dirty rollers Small, circular patterns are produced
8
Hypo Retention Yellowish stain that appears on finished radiograph
Due to inadequate washing Remaining thiosulfate from fixer solution
9
Handling And Storage Artifacts
They occur during darkroom handling or during storage before use. Light or Radiation Fog White light streaks Safelight too close or wrong wattage Film left in x-ray room during exposure
10
Kink Marks Improper handling or storage Appears as a fingernail mark
11
Static Caused by buildup of electrons in the emulsion
Three types of static artifacts Crown Tree Smudge
12
Processing Artifacts
13
FILMS STUCK TO EACH OTHER DURING PROCESSING
14
HYPORETENTION
15
ROLLER MARKS
16
PI-LINE
17
GUIDE SHOE MARK
18
WATER STAIN
19
Handling Artifacts
20
STATIC (Tree Static)
21
STATIC (Crown Static)
22
STATIC (Smudge Static
23
FINGER MARKS
24
CRIMPING MARK
25
CRIMPING MARK
26
PICK-OFF
27
WATER STAIN
28
WATER STAIN
29
Exposure Artifacts
30
MOTION
31
POOR FILM SCREN CONTACT
32
ARTIFACT????
33
H.T. (high Voltage) CABLES
34
ROTATION + FOG
35
DOUBLE EXPOSURE
36
BACKSCATTER OR CASSETTE UPSIDE-DOWN
37
BACKSCATTER
38
BACKSCATTER OR CASSETTE UPSIDE-DOWN
39
MOIRE EFFECT Are artifacts also called zebra pattern Artifact
It is a double set of grid lines caused by the placement of a grid cassette in a Bucky tray.
40
WARPED CASSETTE-LIGHT LEAK
41
HAIR BUN
42
CLOTHES RIBBING
43
TUBE
44
HAIR STREAKING
45
GRID CUT-OFF
46
JEWELLERY
47
HEARING AID
48
LIGHTER IN THE POCKET
49
SCATTER FOG
50
SANDBAG
51
UMBILICAL RING
52
AIR IN THE VENTRICLES AIRHEAD?????
53
PLASTER
54
DIRT ON THE SCREEN
55
PATHOLOGY
56
FOGGING
57
KYPHOSIS-CHIN
58
STRAP
59
SCRATCHES
60
STATIC
61
In CT Scan Artifacts
62
Intra Uterine Device
63
Chole clips
64
Earring Artifact
65
MRI artifacts
66
Patient Movements