Compact Home Symptoms of the problem (click on the relevant line) No display on the unit The machine has frozen up The oxygen consumption is too high. The machine switches on but nothing happens even if you override manually. The right hand pressure gauge is below 2.5 bars. The right hand pressure gauge is above 2.5 bars The pressure on the right hand pressure gauge does not reach 2.5 bars The pressure on the left hand pressure gauge remains at 0. The pressure on the left hand pressure gauge is above 2 bars. The Norton tube (noodle) on the equipment outlet disconnects constantly. The equipment will not run in automatic mode. The equipment does not work and the machine is displaying an error code.
3
Yes No Home Is the equipment connected to a power supply?
4
YesNo Home Disconnect and reconnect the equipment to reinitialise it. Has the machine restarted?
5
YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)
6
YesNo Home Is the pressure on the right hand pressure gauge at 2.5 bars?
7
YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)
8
Manually override in small increments by turning the regulator clockwise to 2.5 bars. Did this help to solve your problem? YesNo Home
9
YesNo Home Is the pressure on the left hand pressure gauge over 2.5 bars?
10
Completely unscrew the regulator, carry out a manual override in order to remove any impurities, then reset the pressure to 2.5 bars. If the problem persists, the doser module is defective. Contact our after- sales service. Did this help to solve your problem? YesNo Home
11
YesNo Home Carry out a manual override. Are the solenoid valves working?
12
YesNo Home Are the Norton tubes (noodles) at the doser outlet and the tank inlet disconnected?
13
If yesIf no Home Check the ceramic element; it may be broken or the pipe connecting to the diffuser may be disconnected.
14
YesNo Home Is the device's outlet coupling connected properly? (risk of a poor connection due to oxygenation).
15
YesNo Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas should flow. Does the left hand pressure gauge fall again immediately?
16
YesNo Home Is the diffuser clogged?
17
If yesIf no Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser and carry out a manual override. The gas should flow and the left hand pressure gauge should fall again immediately.
18
YesNo Home Have the tank and dose volumes been programmed?
19
Yes Home Is the programmed dose compatible with the volume programmed? (see table below) Tank vol 10100250500100020005000 Cali- bration value min - max 50 0.4 - 2000.1 - 900.1 - 400.1 - 20.1 - 90.1 - 4,50.1 - 1.8 100 0.7 - 200 0.1 - 1800.1 - 720.1 - 360.1 - 180.1 - 90.1 - 3.6 250 1.6 - 200 0.2 - 200 0.1 - 1800.1 - 900.1 - 450.1 - 22.50.1 - 9 500 3.2 - 200 0.3 - 200 0.2 - 200 0.1 - 1800.1 - 900.1 - 450.1 - 18 1000 6.3 - 200 0.7 - 200 0.3 - 200 0.2 - 200 0.1 - 1800.1 - 900.1 - 36 2000 12.5 - 200 1.3 - 200 0.5 - 200 0.3 - 200 1.3 - 2000.1 - 1800.1 - 72 500031.3 - 200 3.2 – 200 1.3 - 200 0.7 - 200 0.4 - 200 0.2 - 2000.1 - 180 No
20
If yesIf no Home Check that there are no leaks in the circuit; see attached procedure.
21
Defective machine or supply Contact our after-sales service. Did this help to solve your problem? YesNo Home
22
The equipment is ready for use. Did this help to solve your problem? YesNo Home
23
The O 2 supply circuit is blocked. Contact our after-sales service. Did this help to solve your problem? YesNo Home
24
The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home
25
Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
26
You can repair it with a stainless steel olive or replace the entire tube. ref: 215TA10MO Did this help to solve your problem ? YesNo Home
27
Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home
28
Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
29
Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
30
Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
31
Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
32
The machine has lost its calibration value; to check it disconnect and reconnect the device. The first screen that appears will indicate the calibration value. If this value is 0, you will need to reprogramme it. Programming the calibration value requires you to open the device. Contact our after-sales service. Did this help to solve your problem? YesNo Home
33
Repair following the attached instructions or contact our after-sales service. Did this help to solve your problem? YesNo Home
34
Check the power supply and connections. Did this help to solve your problem? YesNo Home
35
The machine is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
36
Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home
37
While carrying out a manual override, gently tap on the black screw next to the right hand pressure gauge; a solenoid valve may be stuck. Otherwise, the solenoid valve or the solenoid valve coil is defective. Contact our after-sales service. Did this help to solve your problem? YesNo Home
38
The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Possibly replace the regulator outlet coupling. Ref: 3.1.SD1 Did this help to solve your problem? YesNo Home
39
Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home
40
Dosing module defective; contact our after-sales service to replace it. Did this help to solve your problem? YesNo Home
41
Check the regulator setting by turning it anti-clockwise; if the pressure does not drop, the dosing module is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
43
Defective solenoid valve coil. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
44
Leak in the pipe between the device and the tank. Repair it with a joint or replace it. Ref: 1.6Manchon Did this help to solve your problem? YesNo Home
45
Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
46
YesNo Home Is the ceramic element in the diffuser clogged?
47
Internal circuit in device blocked. Contact after-sales service. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
48
Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
49
Enter a tank volume in hL and a dose in mL/L/month. Did this help to solve your problem? YesNo Home
50
Often it is an inputting error. Enter the desired values again. Did this help to solve your problem? YesNo Home
51
The machine has lost its programme, contact our after-sales service. Did this help to solve your problem? YesNo Home
52
Contact our after-sales service. Did this help to solve your problem? YesNo Home
53
Thank you for using Vivelys' online help. We hope that we have met your expectations, and are available to help if you need any further information. Please do not hesitate to contact us by email or telephone: +33 (0)4 67 85 68 40 sav@vivelys.com Home
54
Do you need more help? Were you unable to solve the problem using the online troubleshooting guide? Please contact our technical support team by email or telephone: +33 (0)4 67 85 68 40 sav@vivelys.com Home