1
Pulse Response of 3 CDM Test Head
Three CDM current sensors Measured using two ea. 1 meter long coax cables System 39 ps Risetime BEI ps Risetime, 18 GHz bandwidth ESDA 33 ps Risetime, 16 GHz ringing June 8, 2012 Jon Barth
2
Time Domain Pulse Response
6 GHz Scope and two 1 Meter Coax Cables
3
3 CDM Test Head Discharge Waveforms
CDM discharge waveforms made on RCDM 3 tester Discharge from Special BEI flat surface JEDEC coins of standard diameters 6 GHz Tektronix scope 37 degrees F Dew Point
4
500 Volt Small BEI Coin BEI ESDA JEDEC
5
Impulse Comparison of 3 CDM Current Sensors
These waveforms are inverted 1 GHz Bandwidth
6
Impulse Comparison 3 GHz bandwidth
7
Impulse Comparison 6 GHz bandwidth
8
ESDA-JEDEC CDM Current Sensors
Equal response at 1 GHz Different at 3 GHz Very different responses at 6 GHz BEI 5A Current sensor Wide bandwidth Flat response Accurate measurements for CDM designers
9
Requirements for a Repeatable Standard
Each current sensor to be measured individually Evaluating current sensors in the CDM system includes many variables The Barth Model 5A current sensor is measured individually