THE NQAD PROVIDES EXCELLENT LINEARITY EVEN AT LOW CONCENTRATIONS THAT ARE UNDETECTABLE BY THE ELSD
Application note # QAN-0085 ™ APPLICATION NOTES CHROMATOGRAMS PARAMETERS NQAD GAS: Air FILTER: 1.3 sec GAIN: 1X EVAPORATION TEMP: 35C ELSD GAS: Nitrogen, 1.5 L/min GAIN: 1 EVAPORATION TEMP: 40C HPLC MOBILE PHASE: ACN: Water with 0.2% TFA FLOW RATE: 1.25 mL/min COLUMN: Alltech Alltima C18, 150 x 4.6 mm, 5u COLUMN TEMP: Ambient INJECTION VOLUME :1 µL ANALYTE: Folic Acid Demonstrates better linearity over a wider dynamic range than ELSD QUANT TECHNOLOGIES, LLC TH Lane NE Blaine, MN, USA Note: This application was run at Quant Technologies on a single ELSD and may not be representative of all ELSD’s. 1