Presentation is loading. Please wait.

Presentation is loading. Please wait.

The Process of TAVR Development

Similar presentations


Presentation on theme: "The Process of TAVR Development"— Presentation transcript:

1 The Process of TAVR Development
Stanton J Rowe CEO NXT Biomedical

2 Disclosure Statement of Financial Interest
I, Stanton Rowe, DO have a financial interest/arrangement or affiliation with one or more organizations that could be perceived as a real or apparent conflict of interest in the context of the subject of this presentation. Edwards Lifesciences shareholder, stock options

3 OK, you and your friends have looked into the future, and believe that you have the next big idea in TAVR…

4 2008 slide on 17 TAVR Technologies 4/18 on the market

5 What Problem Are You Solving vs. Current Offerings

6 Virginia Cardiac Services Quality Initiative database.
(2009 to 2011) (2012 to 2015) Virginia Cardiac Services Quality Initiative database.

7 Areas to Improve Vascular complications Prolonged ventilation
Profile and vascular closure devices Prolonged ventilation General anesthesia, COPD Atrial fibrillation Age related incidence; higher age in TAVR Discharge to facility Multifactoral Durability How would you prove better durability with the least clinical implant duration?

8 Similarities and Dissimilaries Surgical versus TAVR Valves
12/9/15 Similarities and Dissimilaries Surgical versus TAVR Valves Similarities Exposed largely to the same boundary conditions Slight differences in removal of calcium in surgery Expectations of durability are largely the same Surgical valves tend to be placed in younger patients VinV has changed the implications of late valve degeneration TAVR tissues and SHV tissues are similar Flow dynamics requirements are the same Host response to implants are the same (early leaflet thickening)

9 Tissue Validation Fixation Chemistry
Mechanical methods applied during Chemical Fixation Testing techniques to assess Biomechanics Uniaxial Biaxial Tensile stress/strain and bending properties Crimp Characterize the condition of leaflets, skirt, and attachment after loading, tracking, and deployment. In the clinical setting, crimp duration can vary therefore these studies should consider worst case crimp durations. All testing for valve performance should be conducted on devices which have been crimped prior to testing. What is even more critical is that the three leaflets have the same mechanical properties. Historically we have focused on matching tissue thickness within any given valve to reduce the chance that one leaflet would stretch more than the other leaflets reducing the coaption length. In fact the thickness of the leaflets is less important that the elasticity of the leaflets. If one leaflet will stretch more than the other leaflets coaption will suffer.

10 Fatigue Testing: Accelerated Wear and Dynamic Failure Mode
200x106 cycles to simulate five years of valve implant duration Peak closed valve pressure difference of /-0 mmHg for Aortic valves Peak closed valve pressure difference of /-0 mmHg for Mitral valves Pressures differences should be continuously monitored throughout the duration of the test Assessments: Hydrodynamic assessments at intervals throughout the test to determine changes in performance Visually examined every 25 x 106 cycles, or until failure Macroscopic damage: Abrasion, holes, tears, delamination, fraying, coaptation, dehiscence.

11

12 Hydrodynamic Testing Pulsatile Flow Pressure drop and Pulsatile Flow Regurgitation Annulus shape, under-deployment must be accounted for in the assessments of flow performance. Broad range of cardiac conditions expected in the patient population (i.e. varied flow rates, pressures and beat rates) Flow Visualization Assess the flow characteristics of the valve using flow visualization or turbulence measurement techniques to characterize any induced jets, flow stasis, leaflet kinematics which might lead to early deterioration in-vivo. not recognized through standard pulsatile flow testing.

13 Delivery Systems A low delivery catheter profile = <14Fr OD
Atraumatic tracking = optimum tracking and flexibility through the vasculature Cost effective Re-capturability and Re-positionability, for self-expanding frame based technologies Ability to orient the rotational alignment with native commissures (Clocking) Intuitive design = ease of use Clear/smart position signaling Marker-bands / clear visual markers under fluoroscopy Stability during deployment Ergonomic Deployment time < 1min Efficient or minimized loading (into Delivery system) process required for delivery of the device Material stability Shelf life > 2 years Guide-wire compatibility Smart packaging e.g. work with the delivery system for TAV loading

14

15 Animal Studies

16 Regulatory IDE First in Human (FIH) Early Feasibility Study (EFS)
Traditional Feasibility Pivotal Clinical Trial (TAVI vs TAVI) (CE Mark) Pivotal Trial (CE Mark) Design Dossier

17 Conclusion TAVR has changed the treatment paradigm for aortic stenosis
It is a growing attractive space for development However, the window for improving TAVR is narrow, the development time and cost are extensive, and established companies continue to raise the bar The pathway is well established


Download ppt "The Process of TAVR Development"

Similar presentations


Ads by Google