Download presentation
Presentation is loading. Please wait.
Published byBronisław Wiśniewski Modified over 5 years ago
1
Cautery (ELECTRO MAGNETIC INTERFERENCE) & cardiac devices in endoscopy
2
background Increasing aging population Advancements in technology
Implantable cardioverter/defibrillators primary therapeutic modality for patients at risk for sudden cardiac death Endoscopy nurses now have patients with multiple comorbidities undergoing procedures
3
Identification of patients
Cardiac implantable electronic devices (CIED) include pacemakers and cardioverter/defibrillator devices (ICD) Any patient with CIED is identified prior to undergoing endoscopy procedures Special precautions need to be taken Multiple co-morbidities?
4
Important considerations
Medical reason for the device Type of device Date when last interrogated ICDs: has the device fired before? Anti-coagulants?
7
Types of Electrocautery
Monopolar cautery (the ground is the patient) Bipolar cautery (the ground is within the instrument) Argon plasma coagulator (APC) (gas/laser)
8
Electrocautery Electrocautery used during endoscopic procedures can cause CIEDs to inappropriately sense because of electromagnetic interference (EMI) Devices can: Inappropriately fire Inappropriately be inhibited Inappropriately trigger cardioversion or defibrillation
9
Pacemakers Several different types:
100% paced all of the time OR Demand- only needed in certain circumstances (i.e. drop in heart rate) Type of pacemaker patient receives depends on indication for pacemaker Patients with pacemakers should have wallet card identifying type of pacemaker they have
10
100% paced
11
Demand pacemaker- atrial fibrillation
12
Implanted Cardiac Defibrillators
Three functions: Pacemaker Overdrive pacing Defibrillation Treated differently than pacemakers
13
Overdrive pacing
16
Defibrillator function
17
Little research available to guide practice, most evidence anecdotal
Few guidelines and most associated with the OR Guertin D, Osman F, et al (2007) Electromagnetic interference and Implanted and Implanted Cardiac Devices Looked at 41 patients with ICDs 40 patients were (L) pectoral placements, 1 (L) abdominal Devices programmed to detection but tachyarrhythmia therapies off
18
41 patients – 52 endoscopies, 5 upper, 28 lower, 8 both, 2 repeat
18 required therapeutic intervention, of these 10 underwent unipolar electrocautery Post procedure ICD interrogation did not document tachyarrhythmia detection or pacemaker inhibition Limits to the study – no oesophageal intervention and small cohort
19
Our practice Policy was developed with the head of our Electrophysiology Lab Ideally patients are pre-identified Pre-admission assessment also includes a section for this Anti-coagulants?
20
Electrocautery & Pacemakers
Do not need cardiac monitoring - monitor heart rate with SaO2 probe Should have magnet in room but does not have to be on patient If cautery is used, the pacemaker may sense the cautery as a heartbeat and not pace when it should Could result in bradycardia If heart rate slows, place magnet over pacemaker- this puts the pacemaker briefly in asynchronous mode When magnet is removed pacemaker resumes normal function
21
Electrocautery & ICDs More complicated
Patient must be on cardiac monitor for the duration of procedure May require two nurses in procedure room Magnet is placed over the device site during cautery only Must be coordinated with physician Magnet should not be held over the device for longer than 30 seconds
22
Assign an RN with prior cardiac nursing experience
Bipolar cautery is best but if monopolar is used, place grounding pad opposite to ICD site, preferably the thigh Magnet is held over the device for the length of the cautery time- remind the physician less than 30 seconds When magnet is not needed, it should be at least 2 ft. away from the device Newer devices will beep if magnet is held too long- the patient will have to go to cardiac lab to have device interrogated
23
Where do we go from here? More research needed as there are more types of therapy out there Spinal or neuro transmitters Cochlear devices Phrenic nerve stimulators
24
References American Society for Gastrointestinal Endoscopy (2007). Endoscopy in patients with implanted electronic devices. Gastrointestinal Endoscopy, 65(4), Capital District Health Authority (2014). Cardiac Implantable Electronic Device (CIED)- Endoscopic Procedures Policy. Retrieved from pe=6&documentRender.GenericField=&documentRender.Id=49926 Guertin, D., Faheen, O., Ling, T., et al (2007). Electromagnetic interference (EMI) and arrhythmic events in ICD patients undergoing gastrointestinal procedure. Pacing and Clinical Electrophysiology, 30(6), Jacob, S., Panaich, S., Maheshwari, R., et al (2011). Clinical applications of magnets on cardiac rhythm management devices. Europace, 13(9), Tsai, A. (2017). Perioperative management of permanent pacemakers (PPMs) and automatic implantable cardioverter-defibrillators (AICDs). Retrieved from emedicine.Medscape.com/article/
25
Questions
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.