1
14026 week 11 Demo Mike Allocco Soham Chakraborty Leslie Havens
Danielle Koch Andrew Miller Kristeen Yee Stephanie Zambito Kris
2
Agenda Team Status The Testing System The MediResp V Database
Elevator Speech Set Expectations Kris
3
Team Status Kris
4
Team Status Kris
5
The Testing System
6
The Tester Length = 5 ¾” Length = 4” Length = 7 ½” Length = 10 ¾”
Andrew Length = 70” Length = 7” Length = 7 ½” Length = 1 ½”
7
The Tester Length = 1” Length = 3 ¾” Length = 1 ¾” Length = 4 ¼”
Andrew Length = 1 15/16”
8
Summary of Airway Resistance Validation Testing
Andrew
9
Pressure Reading Accuracy
Testing the pressure sensors with a manometer gave a pressure to voltage relationship. Currently there is a large difference in calibration between the two sensors Mike
10
Leak Test/Mass Leak Test Configuration 3
Used dish soap method on the final assembly No leaks were found Set-up and testing videos can be found on the team’s Edge site Mass including mounting= 2.995kg Leslie
11
Problem Tracking Depending on problem
12
The Mediresp V
13
PEV VS Testing SYSTEM ACCURACY
Test ID Time (sec) Maximum Flow Rate (LPM) PEV Display (LPM) PEV Volume % Error 1 10.22 11.6 12 380 3% 2 10.21 13.5 15 11% 3 10.23 17 18 6% 4 20 21 5% 5 11.3 6 7 16.8 19 13% Kris
14
Human analogue - PEV Breath Rate: Tidal Volume:
~ bpm for infants ~ bpm for adults Tidal Volume: Can be expressed as for infants: 18.1 mL/kg for healthy adults: (0.0275* Age[y] *Height[cm] ) L for obese adults: (0.0277* Age [y] *Height[cm] ) L for restrictive diseases (volumes normal, flow rates impeded): (from ) * (normal tidal volume) L for obstructive diseases (volumes are decreased): (0.8 – 1.0) * (normal tidal volume) L MILLER, WAYNE C.; SWENSEN, THOMAS; WALLACE, JANET P. (February 1998). "Derivation of prediction equations for RV in overweight men and women". Medicine & Science in Sports & Exercise 30 (2): 322–327. Danielle
15
Human analogue - tester
Airway Resistance: Poiseuille’s Law of Resistance R ~= 1/r4 where r is the radius of trachea For neonates: 19.2± L/s For adults: cmH2O/L/s Compliance: For neonate: 0.5 to 2.0 mL/cm H2O For adults: ~ 0.1 L/cmH20 Danielle
16
S tESTS 1-5 (of 9) Breath Rate Inspiration Time Rise Time Tidal Volume
Inhale:Exhale Ratio Danielle
17
BREAKING DOWN WAVEFORM
Breath Rate: 1 breath in 2.02 sec 29.7 bpm Rise Time 0.50 sec Insp. Time 1.00 sec Pause 1.04 sec Inhale:Exhale = Insp. Time/Pause 0.96:1 (I:E) Tidal Volume 0.72 L Rise Time Danielle Inspiratory Time Pause At nominal Q=12 Lpm, V=700 mL
18
Pressure Provided/Mass
Pressure Provided by the PEV We are reading negative pressures before the orifice plate and positive pressure after Mass of PEV Our results = kg Team 13027’s Result = 4.2 kg Leslie
19
PEV-Human Connection Performed a leak test using dish soap.
No bubbles, so no leak Assumed the mask against the patient's face was secure, since it is a mask for medical purposes. Leslie
20
Open Items Extrinsic Peak End-Expiratory Pressure Mean Airway Pressure
Intrinsic peak End-expiratory Pressure Danielle
21
Template for Database - Summary
Kris
22
Template for Database - Summary
Kris
23
Template for Database - Data
Kris
24
Elevator Speech A portable emergency ventilator provides air to a person who is incapable of breathing on his or her own. Two previous senior design teams developed the latest versions of the MEDIRESP which is a unique ventilator that has a CPR mode. Before the MEDIRESP can go to market it must be tested. Our team’s project is to characterize the capabilities of the latest MEDIRESP design. Our system models the human respiratory system in order to collect flow, pressure, and volume data during different MEDIRESP settings. From this information we can determine if the MEDIRESP would be successful in an emergency setting. Steph
25
Team Vision Weeks 12-14 At the end of MSD II phase 5 the team will have a demonstration of our project’s progress. Our expectations for our demonstration are: Complete all S test plans. Be able to show the customer our test results for tests all S tests. Be prepared to describe how we tested each requirement. Be prepared to show how we tested each requirement. Prepare and run our Imagine RIT exhibit. See Imagine RIT Vision for more details. Complete project technical paper. Complete project poster. Submit application for poster contest by May 4. Kris
26
Go to Gemba!
27
Go to Gemba
28
Questions?