CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON

Similar presentations


Presentation on theme: "CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON"— Presentation transcript:

1 CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON
LT Scott Boedeker

2

3

4 CTD 10/XBT 4

5 REPRESENTATIVE PAIRS

6

7 For the 27 comparisons from this cruise, the XBT temperature has a warm bias of degrees C over the depth range This is similar to results obtained from previous OC3570 cruises The greatest bias was in the upper levels

8 CTD-XBT Temp Difference
R/V Point Sur 08/ / /2000 Roth (2001) Schmeiser (2000) Depth (m) Mean (C) Std 25-125 0.5135 0.1779 0.3598 0.2131 0.0951 0.1975 0.2157 0.0960 0.1981 0.2372 0.0903 0.2194 Sfc-750 0.2147 0.1047 0.2151

9 Why the Bias? Manufacturing Defects Method of deployment
-CTD casts are made nearly vertical with the vessel stopped. -XBT casts are made with the vessel in motion and are therefore not as vertical. This results in the XBT being shallower than it is recorded.

10 Questions?