1
Instruments used HORIBA AIA721 NDIR online analyzer
integrated in emission bench Varian GC 450 with ECD offline analysis total run time 7 min WLTP-DTP-AP
2
Background measurement from air bag
GC shows good repeatabiliy with low scatter (Std = 1,6 ppb) NDIR shows offset of -35 ppb and significantly higher scatter (Std = 29 ppb) WLTP-DTP-AP
3
Measurement from sample bag
Deviation of NDIR to GC of up to 25 % Interference correction has large influence on NDIR results WLTP-DTP-AP
4
Emission results from FTP75 tests
WLTP-DTP-AP
5
Simulated emission results at low level
Sample bags were assumed with 0,5 ppm for all tests WLTP-DTP-AP