1 Isolated Bad-CAC Strip at UCSB
For first 57 modules produced at UCSB, there was a slight misunderstanding of bonding rules Bonded isolated Bad CAC channel that were not also marked as a bad IDIEL channel 23 of 57 modules had at least one such channel 42 channels total Plucking channels would result in a few more Grade B modules Not a large concern

2 Testing Results on Isolated Bad CAC Strips
100 Hz measured (in pF) at QTC All modules have been fully tested with ARCS and LT system Both pulsed and continuous LED tests taken ALL of the 42 strips had regular test results (passed) No problem seen with these channels Yes, that’s negative capacitance!!!