Presentation is loading. Please wait.

Presentation is loading. Please wait.

Tungsten Cable Failure Analysis Brad James Ph.D. P.E.

Similar presentations


Presentation on theme: "Tungsten Cable Failure Analysis Brad James Ph.D. P.E."— Presentation transcript:

1 Tungsten Cable Failure Analysis Brad James Ph.D. P.E.

2 Tungsten Cable Failure Analysis Background Medical device company uses tungsten cables as part of a surgical apparatus Non-sterilized cables withstand hundreds of durability test cycles (consists of cyclic and static loading) Sterilized cables fail after a few durability test cycles Tensile testing shows no real difference between sterilized and non-sterilized cables

3 Tungsten Cable Failure Analysis

4 SterilizedNon-Sterilized

5 Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing

6 Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing

7 Tungsten Cable Failure Analysis Cables After Tensile Testing: Sterilized Non-Sterilized

8 Tungsten Cable Failure Analysis Hydrogen content tests (Leco Interstitial) Non-sterilized cable: ~20 ppm hydrogen Sterilized cable: ~ 60 ppm hydrogen

9 Tungsten Cable Failure Analysis Conclusions Classic case of hydrogen embrittlement! (tungsten is known to be susceptible) Corrosion during sterilization is the source of hydrogen Embrittlement is stress/time dependent Tensile testing not long enough to allow diffusion of hydrogen to high stress locations Static loads during durability allowed hydrogen to diffuse to higher stress locations – resulting in fracture


Download ppt "Tungsten Cable Failure Analysis Brad James Ph.D. P.E."

Similar presentations


Ads by Google