Download presentation
Presentation is loading. Please wait.
Published byMarvin O'Neil Modified over 10 years ago
1
Tungsten Cable Failure Analysis Brad James Ph.D. P.E.
2
Tungsten Cable Failure Analysis Background Medical device company uses tungsten cables as part of a surgical apparatus Non-sterilized cables withstand hundreds of durability test cycles (consists of cyclic and static loading) Sterilized cables fail after a few durability test cycles Tensile testing shows no real difference between sterilized and non-sterilized cables
3
Tungsten Cable Failure Analysis
4
SterilizedNon-Sterilized
5
Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing
6
Tungsten Cable Failure Analysis Sterilized Cable After Durability Testing
7
Tungsten Cable Failure Analysis Cables After Tensile Testing: Sterilized Non-Sterilized
8
Tungsten Cable Failure Analysis Hydrogen content tests (Leco Interstitial) Non-sterilized cable: ~20 ppm hydrogen Sterilized cable: ~ 60 ppm hydrogen
9
Tungsten Cable Failure Analysis Conclusions Classic case of hydrogen embrittlement! (tungsten is known to be susceptible) Corrosion during sterilization is the source of hydrogen Embrittlement is stress/time dependent Tensile testing not long enough to allow diffusion of hydrogen to high stress locations Static loads during durability allowed hydrogen to diffuse to higher stress locations – resulting in fracture
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.