Download presentation
1
Electromyography: Recording
D. Gordon E. Robertson, PhD, FCSB Biomechanics Laboratory, School of Human Kinetics, University of Ottawa, Ottawa, Canada
2
Biomechanics Laboratory, University of Ottawa
EMG Recording: Topics Surface or indwelling Electrode placement Type of amplifier Common Mode Rejection Ratio (CMRR) Dynamic range and Gain Input impedance and skin resistance Frequency response Telemetry versus directly wired Biomechanics Laboratory, University of Ottawa
3
Biomechanics Laboratory, University of Ottawa
Types of Electrodes Bipolar surface Needle Fine-wire Biomechanics Laboratory, University of Ottawa
4
Biomechanics Laboratory, University of Ottawa
Surface Electrodes lower frequency spectrum (20 to 500 Hz) relatively noninvasive, cabling does encumber subject, telemetry helps skin preparation usually necessary surface muscles only global pickup (whole muscle) inexpensive and easy to apply Biomechanics Laboratory, University of Ottawa
5
Biomechanics Laboratory, University of Ottawa
Surface Electrodes pre-gelled disposable electrodes are most common and inexpensive MLS pre-amplified electrodes reduce movement artifact Delsys Trigno includes 3D accelerometers Biomechanics Laboratory, University of Ottawa
6
Indwelling Electrodes
fine wire or needle localized pickup difficult to insert invasive, possible nerve injury produces higher frequency spectrum (10 to 2000 Hz) can record deep muscles Biomechanics Laboratory, University of Ottawa
7
Biomechanics Laboratory, University of Ottawa
Electrode Placement electrode pairs in parallel with fibres midway between motor point and myotendinous junction (or near belly of muscle) approximately 2 cm apart, better if electrodes are fixed together to reduce relative movement Biomechanics Laboratory, University of Ottawa
8
Surface Electrode Placement
myotendinous junctions frequency spectra motor point best strongest EMG Biomechanics Laboratory, University of Ottawa
9
Noise Reduction and Grounding
leads should be immobilized to skin surgical webbing can help reduce movement artifacts ground electrode placed over electrically neutral area usually bone N.B. there should be only one ground electrode per person to prevent “ground loops” that could cause an electrical shock Biomechanics Laboratory, University of Ottawa
10
Surface Electrode System (preamplifier type)
Differential amplifier Ground or reference electrode Cable Leads Electrodes Biomechanics Laboratory, University of Ottawa
11
Biomechanics Laboratory, University of Ottawa
Type of Amplifier because EMG signals are small (< 5 mV) and external signals (radio, electrical cables, fluorescent lighting, television, etc.) are relatively large, EMG signals cannot be distinguished from background noise background noise (hum) is a “common mode signal” (i.e., arrives at all electrodes simultaneously) common mode signals can be removed by differential amplifiers single-ended (SE) amplifiers may be used after differential preamplified electrodes Biomechanics Laboratory, University of Ottawa
12
Common Mode Rejection Ratio (CMRR)
ability of a differential amplifier to perform accurate subtractions (attenuate common mode noise) usually measured in decibels (y = 20 log10 x) EMG amplifiers should be >80 dB (i.e., S/N of :1, the difference between two identical 1 mV sine waves would be 0.1 mV) most modern EMG amplifiers are >100 dB Biomechanics Laboratory, University of Ottawa
13
Biomechanics Laboratory, University of Ottawa
Dynamic Range and Gain dynamic range is the range of linear amplification of an electrical device typical A/D computers use +/–10 V or +/–5 V amplifiers usually have +/–10 V or more, oscilloscopes and multimeters +/–200 V or more audio tape or minidisk recorders have +/–1.25 V EMG signals must be amplified by usually 1000x or more but not too high to cause amplifier “saturation” (signal overload) if too low, numerical resolution will comprised (too few significant digits) Biomechanics Laboratory, University of Ottawa
14
Biomechanics Laboratory, University of Ottawa
Input Impedance impedance is the combination of electrical resistance and capacitance all devices must have a high input impedance to prevent “loading” of the input signal if loading occurs the signal strength is reduced typically amplifiers have a 1 MW (megohm) input resistance, EMG amplifiers need 10 MW or greater 10 GW bioamplifiers need no skin preparation Biomechanics Laboratory, University of Ottawa
15
Biomechanics Laboratory, University of Ottawa
Skin Impedance dry skin provides insulation from static electricity, 9-V battery discharge, etc. unprepared skin resistance can be 2 MW or greater except when wet or “sweaty” if using electrodes with < 1 GW input resistances, skin resistance should be reduced to < 100 kW Vinput = [ Rinput / (Rinput + Rskin) ] VEMG Biomechanics Laboratory, University of Ottawa
16
Skin Impedance: Example
Vinput = [ Rinput / (Rinput + Rskin) ] VEMG If skin resistance is 2 MW (megohm) and input resistance is 10 MW then voltage at amplifier will be [10/(10 + 2) = 0.833] 83.3% of its true value. By reducing skin resistance to 100 kW this can be improved to 99%. By also using a 100 MW resistance amplifier the signal will be 99.9%. Biomechanics Laboratory, University of Ottawa
17
Biomechanics Laboratory, University of Ottawa
Frequency Response frequency responses of amplifier and recording systems must match frequency spectrum of the EMG signal since “raw” surface EMGs have a frequency spectrum from 20 to 500 Hz, amplifiers and recording systems must have same frequency response or wider since relative movements of electrodes cause low frequency “artifacts,” high-pass filtering is necessary (10 to 20 Hz cutoff) since surface EMG signals only have frequencies as high as 500 Hz, low-pass filtering is desirable (500 to 1000 Hz cutoff) therefore use a “band-pass filter” (e.g., 20 to 500 Hz) Biomechanics Laboratory, University of Ottawa
18
Biomechanics Laboratory, University of Ottawa
Frequency Response Typical frequency spectrum of surface EMG Biomechanics Laboratory, University of Ottawa
19
Biomechanics Laboratory, University of Ottawa
Typical Band Widths EMG 20–500 Hz 10–1000 Hz surface indwelling ECG 0.05–30 Hz 0.05–100 Hz standard diagnostic EEG 1–3 Hz 4–7 Hz 8–12 Hz 12–30 Hz 30–100 Hz delta waves theta waves alpha waves beta waves gamma waves muscle forces or human movements DC–10 Hz muscle moments joint trajectories audio 20–8000 Hz 20– Hz 20– Hz voice tape CD Biomechanics Laboratory, University of Ottawa
20
Biomechanics Laboratory, University of Ottawa
EMG Sampling Rate since highest frequency in surface EMG signal is 500 Hz, A/D (computer) sampling rates should be 1000 Hz or greater (>2 times maximum frequency) raw EMGs cannot be correctly recorded by pen recorders since pen recorders are essentially 50 Hz low-pass filters mean or median frequencies of unfatigued muscles are around 70 to 80 Hz “notch” filters should not be used to remove 50/60 cycle (line frequency) interference because much of the EMG signal strength is in this range Biomechanics Laboratory, University of Ottawa
21
Telemetry versus Direct Wire
telemetry has less encumbrance and permits greater movement volumes radio telemetry can be affected by interference and external radio sources radio telemetry may have limited range due to legislation (e.g., IC, FCC, CRTC) cable telemetry (e.g., Bortec) can reduce interference from electrical sources telemetry is usually more expensive than directly wired systems telemetry has limited bandwidth (more channels reduce frequency bandwidths) Biomechanics Laboratory, University of Ottawa
22
Biomechanics Laboratory, University of Ottawa
Telemetered EMG Delsys’s Trigno EMG and accelerometry telemetry system Biomechanics Laboratory, University of Ottawa
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.