Intrinsic Uniformity QC Setup
Sunday Case of the Day Physics Patricia Svolos1, Matt Vanderhoek2 and Beth Harkness2 1McGovern Medical School, University of Texas Health Science Center at Houston, TX 2Henry Ford Health System, Detroit, MI Detector 1 Detector 2 Point Source Intrinsic Uniformity QC Setup PASS FAIL Extrinsic Uniformity Intrinsic History: Daily QC (Extrinsic Uniformity) of a dual head SPECT/CT system using a 153Gd rod source passes (uniformity ≤ 5%). The weekly QC (Intrinsic Uniformity) is performed on the same day using a 99mTc point source and fails for Detector 1 (uniformity ≥ 4%). What is the cause of the linear dark band artifact present on the left side of the Intrinsic Uniformity QC image for Detector 1? The collimator on Detector 1 was contaminated with Tc-99m. Bad linearity correction resulting in a linear band at the left side of the image. External source of radiation being partially blocked by the edge of the gamma camera. Band of defective photomultiplier tubes on the left side of the camera crystal.
2
Diagnosis: C. External source of radiation being partially blocked by the edge of the
camera crystal. Intrinsic QC is performed without the collimators on the system. Contaminated waste in the trash was in the room during QC testing. The 180o orientation of the camera heads was such that Detector 1 faced down toward the floor while Detector 2 faced up toward the ceiling. Consequently, the contaminated trash (on the floor) irradiated the face of Detector 1 (yellow lines) but not the face of Detector 2. Photons were attenuated at certain angles by the leaded light rail on Detector 1, thus creating a linear “shadow” of the light rail on the edge of the acquired uniformity image. Detector 1 Detector 2 Detector 1 Detector 2 Light rail
3
Discussion: A. The collimator on Detector 1 was contaminated with Tc-99m
Incorrect Collimators are used only during extrinsic uniformity QC. Collimator contamination of Detector 1 with Tc-99m would likely result in an artifact on the extrinsic uniformity QC image with a random pattern (looking more “splotchy”) rather than a linear dark band artifact. The acquired extrinsic uniformity image for Detector 1 was free of such artifacts.
4
Discussion: B. Bad linearity correction resulting in a linear band at the left side of the image
Incorrect Artifacts due to bad linearity correction usually have a “tubular” pattern rather than linear. Image below shows a typical artifact pattern of all PMT being visible if linearity correction fails.
5
Discussion: D. Band of defective photomultiplier tubes on the left side of the camera crystal
Incorrect It is unlikely that a set of defective photomultiplier tubes near the edge of the crystal would result in a linear dark band artifact. Image below shows a typical artifact due to bad photomultiplier tubes.
6
References/Bibliography:
1. Physics in Nuclear Medicine. Cherry S.R., Sorenson J.A., Phelps M.E.