1 Synchromed II Programming

2 1) At least 20mins after MRI, turn on the N’Vision and press the ‘pump’ icon on the screen:
2) Choose ‘Synchromed II’ 3) Press ‘Interrogate’ to start communication with the pump

3 6) Position programming head over pump and press ‘OK’
4) Allow full interrogation until telemetry is complete and select ‘ok’ 5) Tick ‘Logs’ from options and press ‘OK’

4 7) Pump status report is displayed with current pump settings
7) Pump status report is displayed with current pump settings. Select ‘X’ to close 8) Select ‘toolki’t icon 9) Select ‘pump logs ‘ tab

5 10) Select ‘Get Logs’ 11) Event log data screen displays all event history including alarms or PA requests