1 Irradiated FBK 3D sensor lab tests
Enver Alagoz1 On behalf of CMS 3D collaboration 1Physics Department, Purdue University, West Lafayette, IN US 15 March 2012

2 Introduction Irradiation 800 MeV protons/cm2 at Los Alamos
Fluences: 1E15, and 5E15 p/cm2 No annealing applied All ROCs work after irradiation Lab measurements performed in the thermal chamber Leakage current Noise Source data

3 Sensors p+ and n+ electrodes pass through p+ substrate completely
Guard rings 200 μm thickness ATLAS08 batch (2010)

4 Cold setup Cold setup for irradiated 3D sensor testing
Environmental chamber -35 °C < T < 200 °C Humidity control Works with CAPTAN and PSI DAQ systems Good for sensor IV measurements and ROC calibrations at cold Manual temperature control Purdue cold test setup Sensor ROC Bump bonds VHDI Base plate Adhesive Wire bond Bias wire

5 Cold setup Purdue cold test setup HV Sensor Sr-90 source
Linear actuator Purdue cold test setup Temperature sensor

6 Thermal measurements in the chamber
Infrared camera placed in the chamber to measure the sensor temperature Thermal chamber at 10 °C Thermal image of sensor at emissivity = 1 Purdue cold test setup

7 Thermal measurements in the chamber
Thermal chamber at 10 °C ROC is ON pcb ΔT = 13 °C --- temperature sensor ROC is OFF

8 ROC tests Sensor: 1E_1_W8 Φ = 5E15 p/cm2

9 IV measurements Thermal chamber at -20 °C

10 Noise tests Thermal chamber at -20 °C

11 Source testing Thermal chamber at -20 °C
Sr-90 source: 1 mCi, Eβ = ΜeV Random trigger used Thermal chamber at -20 °C -15V Φ = 1E15 p/cm2 1 Vcal = 65 e- MP = 10.6 ke- (Thickness: 200 μm)

12 Source charge collection scan
Sr-90 source: 1 mCi, Eβ = ΜeV Random trigger used Thermal chamber at -20 °C Landau convoluted Gaussian fit -15V Φ = 1E15 p/cm2

13 Charge collection scans
Thermal chamber at -20 °C

14 Source testing: pre- and post-irradiation

15 Summary All irradiated FBK sensors work
All ROCs works after irradiation Lab measurements performed in the thermal chamber at -20 °C Thermal measurements suggest sensor temperature is -7 °C (ROC ON) when thermal chamber at -20 °C due to ΔT = 13 °C IV, noise scan, and source testing done Negligible increase in noise Leakage current increased Few volts increase in depletion voltage Signal LOSS 1E 43% after 1E15 p/cm2 1E 50% after 5E15 p/cm2 2E 14% after 1E15 p/cm2 4E 14% after 1E15 p/cm2