New 3-Dimensional Scheimpflug Analysis of Cataract Density
International Vision Correction Research Centre New 3-Dimensional Scheimpflug Analysis of Cataract Density T.M. Rabsilber, MD, S. Hara, MD, I.J. Limberger, MD, A. Ehmer, MD, G.U. Auffarth, MD Dept. of Ophthalmology, University of Heidelberg, Germany, Director: Prof. Dr. H.E. Völcker The authors have no financial interest in the products mentioned here
2
Introduction The subjective grading of the density of cataract is…
- time-consuming - complicated - less suitable for longitudinal trial An objective grading is required
3
Introduction Pentacam® (Oculus, Germany) is an instrument for imaging the anterior segment of the eye by a rotating Scheimpflug camera measurement.
4
Introduction With Pentacam® we can grade the density of cataract by quantifying the density and the area of the lens opacification in 3-dimension with the specific software installed.
5
Purpose We objectively and 3-dimentionally graded the density of cataract into 5 grades using Pentacam®. We investigated the difference in the mean total phaco energy during phacoemulsification comparing the 5 grades.
6
Patients and Methods 186 eyes of 166 patients undergoing phacoemulsification for cataract using Infinity® (Alcon Inc.) by only one surgeon. They were preoperatively examined with Pentacam® and graded into 5 grades according to the density and the area of the opacification of lenses. There is no case with any complication neither during nor after surgery.
7
Patients and Methods The phaco time and the average phaco power were examined and the total phaco energy was calculated for each case. Kruskal-Wallis Test was used to analyze statistically the difference in the mean total phaco energy between 5 groups.
8
Grading with Pentacam®
Grade 1 Grade 2 Grade 3 Grade Grade 5 These photographs show the images of each grade. As the grade increases, a higher density of the lens center is recognized.
9
Numbers in each grade
10
The grade of the density of cataract and the mean total phaco energy
11
Results The mean phaco energies were:
The mean phaco energies were increasing with higher grades. In the group of the G1, the mean phaco energy was statistically lower than any other groups (from p<0.001 to p<0.05).
12
Conclusion Mean phaco energy increased with higher density grades.
It is possible to use Pentacam® as a new objective 3-D grading system of the cataract density.