New 3-Dimensional Scheimpflug Analysis of Cataract Density

Similar presentations


Presentation on theme: "New 3-Dimensional Scheimpflug Analysis of Cataract Density"— Presentation transcript:

1 New 3-Dimensional Scheimpflug Analysis of Cataract Density
International Vision Correction Research Centre New 3-Dimensional Scheimpflug Analysis of Cataract Density T.M. Rabsilber, MD, S. Hara, MD, I.J. Limberger, MD, A. Ehmer, MD, G.U. Auffarth, MD Dept. of Ophthalmology, University of Heidelberg, Germany, Director: Prof. Dr. H.E. Völcker The authors have no financial interest in the products mentioned here

2 Introduction The subjective grading of the density of cataract is…
- time-consuming - complicated - less suitable for longitudinal trial An objective grading is required

3 Introduction Pentacam® (Oculus, Germany) is an instrument for imaging the anterior segment of the eye by a rotating Scheimpflug camera measurement.

4 Introduction With Pentacam® we can grade the density of cataract by quantifying the density and the area of the lens opacification in 3-dimension with the specific software installed.

5 Purpose We objectively and 3-dimentionally graded the density of cataract into 5 grades using Pentacam®. We investigated the difference in the mean total phaco energy during phacoemulsification comparing the 5 grades.

6 Patients and Methods 186 eyes of 166 patients undergoing phacoemulsification for cataract using Infinity® (Alcon Inc.) by only one surgeon. They were preoperatively examined with Pentacam® and graded into 5 grades according to the density and the area of the opacification of lenses. There is no case with any complication neither during nor after surgery.

7 Patients and Methods The phaco time and the average phaco power were examined and the total phaco energy was calculated for each case. Kruskal-Wallis Test was used to analyze statistically the difference in the mean total phaco energy between 5 groups.

8 Grading with Pentacam®
Grade 1 Grade 2 Grade 3 Grade Grade 5 These photographs show the images of each grade. As the grade increases, a higher density of the lens center is recognized.

9 Numbers in each grade

10 The grade of the density of cataract and the mean total phaco energy

11 Results The mean phaco energies were:
The mean phaco energies were increasing with higher grades. In the group of the G1, the mean phaco energy was statistically lower than any other groups (from p<0.001 to p<0.05).

12 Conclusion Mean phaco energy increased with higher density grades.
It is possible to use Pentacam® as a new objective 3-D grading system of the cataract density.