Lunar observation plan based on SI-traceable demonstration instrument development Feng Jiang Shanghai Institute of Technical Physics , Chinese Academy of Science 2017 Joint GSICS/IVOS Lunar Calibration Workshop
2
2017 Joint GSICS/IVOS Lunar Calibration Workshop
1.Lunar observation plan introduction 2.Sensor calibration overview 3.Error budget analysis 4.On orbit calibration
3
Orbit parameters Moon Sun Spectrometer Altitude ~ 410Km Inclination
50° Orbit period 92.7 minutes Ground speed 7.23Km/s Spectrometer
4
+X +Y +Z
5
XOY plane -90° -60° -30° 0° 30° 60° 90°
7
Cryogenic Absolute Radiometer
Spectral range: 0.25~50μm Absolute accuracy: <0.02% Power resolution: 0.01nW Key Laboratory of Optics Calibration and Characterization, Chinese Academy of Science
8
Trap Detector
9
Spectral reponsivity
10
spatial non-uniformity
instability polarization sensitivity repeatability linearity laser power measurement
14
Absolue spectral response uncertainty of standard detector @632.8nm
15
Standard radiometer
16
Standard radiometer
17
Standard radiometer
19
Spectrometer calibration uncertainty
-Uncertainty of the standard radiometer -Stability and spatial uniformity of the light source -Wavelength calibration -Polarization sensitivity -Stray light -Detector linearity -Point spread function and line spread function -Thermal
20
Pre-flight calibration
Integrating Sphere Spectrometer Tunable Laser Standard radiometer
21
Polarization sensitivity
zigzag blazed grating
22
Birefringence index difference
Depolarizer Clear Aperture 40mm Wedge angle 12'±2' Center thickness t1 /mm 3.5±0.01 Center thickness t2 /mm 3.53±0.01 Material quartz crystal Birefringence index difference 0.009
23
Cryogenic Absolute radiometer Spectrometer (VIS~NIR)
Error Budget Cryogenic Absolute radiometer Uncertainty ~0.02% Si trap detector Uncertainty ~0.15% Spectrometer (VIS~NIR) Uncertainty ~1.24% Standard Radiometer Uncertainty ~0.5% 激光均匀光源 稳定性 (Laser illuminated integrating sphere source)
24
On orbit calibration
25
On orbit calibration
26
On orbit calibration On orbit attenuator Ground
(Neutral density filter)
27
Thank you !