Download presentation
Presentation is loading. Please wait.
1
Introduction to Flow Cytometry
IGC Workshop Fundamentals of Flow Cytometry (cont.) Rui Gardner IGC – April 02, 2013
2
The Instrument 2
3
Optics (Emission Detectors)
4
Filters LP : Long Pass Filter > 500 BP : Band Pass Filter 530 / 60
Filtered Blocked LP : Long Pass Filter > 500 Filtered Blocked BP : Band Pass Filter 530 / 60 4
5
Filters 5
6
Optical Layout 6
7
Detectors and Signal Processing
8
Detection PMT Photo Multiplier Tube
PMT’s collect photons that are then converted into voltage signals 8
9
Pulse Flowing Stream Voltage pulse Laser 9
10
Pulse Parameters W H A H: A: W: 10
11
Height vs Area H H A A For non spherical cells, Height (FL-H) is not an adequate parameter to analyze Area (FL-A) is the most adequate. However, we still need to remove doublets from the analysis... 11
12
Doublet Discrimination
W 2W H W 2H H A 2A 2A FL-W FL-A Single cells doublets FL-H 12
13
Threshold Forward Scatter Threshold Voltage H Threshold Time W
Small Cells and debris Cells of Interest 13
14
Data Handling
15
Histograms Cell Count Fluorescence in a cell Fluorescence
16
What are those dots?
17
Gating Common Gate Shapes Logical Gating AND, OR, NOT 17
18
Gating Positive or Negative?
A “positive” cell or event is that which falls outside the “negative” gate. Neg Pos 18
19
Back Gating Back gating a positive population can enrich the population of interest and help identify it correctly CD4 FITC Lymphocytes CD4- CD4+ 19
20
Dot vs Countour Plots Dot Plots Contour and Density Plots 20
21
Logarithmic or Linear? Anti-CD4-labeled antibody
Signals vary >100-fold Use Log scale Linear Log DNA-labeling dye Signals vary 2- to 10-fold Use Linear scale Linear Log 21
22
Acquisition How many cells should I acquire? Precision cv % = sd mean
x 100 Counting cells follows Poisson statistics: cv % = 1 x 100 N is the number of cells counted Example: Population of interest is 1% of total population and want 5% precision N = 1002 (cv %)2 = 10,000 25 = 400 40,000 Number of cells to be counted in the region of interest Number of total cells to be counted 22
23
Analysis Software FlowJo X VenturiOne CellQuest Summit FCSExpress
Kaluza FACSDiva 23
24
Analysis Software Commercial Analysis Software:
Open Source or Free software: FlowJo X (TreeStar) - Site IGC CellQuest/Pro (BD) - 3 IGC FACSDiva (BD) - 2 IGC VenturiOne (Applied Cytometry) FCS Express (De Novo Software) GemStone (Verity Software) WinList (Verity Software) ModFit LT (Verity Software) FlowLogic (Inivai) Kaluza (Beckman Coulter) Guava Software (Merck-Millipore) Summit (Beckman Coulter) FlowCore (R/Bioconductor) Weasel Flowing Software Attune Software (Life Technologies) FCS Express (De Novo Software) Cyflogic Plate Analyzer Softwares Plate Analyzer (Free – Purdue University) Hyperview (Commercial – Intellicyt) FACSDiva (Commercial – BD) 24
25
Introduction to Flow Cytometry
IGC Workshop Fundamentals of Flow Cytometry (end) Rui Gardner IGC – April 02, 2013
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.