Next Home Visio2 troubleshooting guide Start the slide show then click Next
2
VISIO2 after 2007 Home Symptoms of the problem (click on the relevant line) No control screen display. The control screen does not switch on or has frozen up. The oxygen consumption is too high. The control screen works but nothing happens, even if you override manually. The right hand pressure gauge is below 2.5 bars on all dosers. The right hand pressure gauge is above 2.5 bars on one of the dosers. The pressure on the right hand pressure gauge does not reach 2.5 bars on one of the dosers. The pressure on the left hand pressure gauge stays at 0 on one of the dosers. The pressure on the left hand pressure gauge is above 2 bars on one of the dosers. The Norton tube (noodle) on the outlet of one of the dosers disconnects constantly. The equipment will not run in automatic mode. The equipment does not work and the machine shows an error code. The right hand pressure gauge is below 2.5 bars on one of the dosers.
3
YesNo Home Is the equipment connected to a power supply?
4
YesNo Home Disconnect and reconnect the device to reinitialise it. Has the machine restarted?
5
YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)
6
YesNo Home Is the pressure on the right hand pressure gauge at 2.5 bars?
7
YesNo Home Is the equipment connected to the regulator properly? (sometimes the connection is poor due to oxygenation)
8
YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)
9
Manually override in small increments by turning the regulator clockwise to 2.5 bars. Did this help to solve your problem? YesNo Home
10
YesNo Home Is the pressure on the left hand pressure gauge over 2.5 bars?
11
Completely unscrew the regulator, carry out a manual override in order to remove any impurities, then reset the pressure to 2.5 bars. If the problem persists, the doser module is defective. Contact our after- sales service. Did this help to solve your problem? YesNo Home
12
YesNo Home Carry out a manual override. Are the solenoid valves working?
13
YesNo Home Are the Norton tubes (noodles) on the doser outlet and the tank inlet disconnected?
14
If yesIf no Home Check the ceramic element; it may be broken or the pipe connecting to the diffuser may be disconnected.
15
YesNo Home Is the device's outlet coupling properly connected? (risk of a poor connection due to oxygenation).
16
YesNo Home Is the diffuser clogged?
17
YesNo Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas should flow. Does the left hand pressure gauge fall again immediately?
18
YesNo Home Is the diffuser clogged?
19
YesNo Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas should flow. Does the left hand pressure gauge fall again immediately?
20
YesNo Home Have the tank and dose volumes been programmed?
21
Yes Home Is the programmed dose compatible with the volume programmed? (max 120mL/L/month). No
22
If yesIf no Home Check that there are no leaks in the circuit (see attached procedure).
23
Defective machine or supply Contact our after-sales service. Did this help to solve your problem? YesNo Home
24
The equipment is ready for use. Did this help to solve your problem? YesNo Home
25
The O 2 supply circuit is blocked. Contact our after-sales service. Did this help to solve your problem? YesNo Home
26
The O 2 supply circuit is blocked. Contact our after-sales service. Did this help to solve your problem? YesNo Home
27
The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home
28
Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
29
You can repair it with a stainless steel olive or replace the entire tube. ref: 215TA10MO Did this help to solve your problem ? YesNo Home
30
Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home
31
Clean the ceramic element (see attached procedure). Check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
32
Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
33
Clean the ceramic element (see attached procedure). Check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home
34
Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
35
The machine has lost its calibration value; you can check it on the "SETTINGS" screen. If this value is 0, you will need to reprogramme it. The machine is programmed in the "SETTINGS" menu: enter the code "8574" then enter the calibration value for each doser and reset the code to 0. Did this help to solve your problem ? YesNo Home 0.0 USER CODE SYSTEM CODE
36
Repair following the attached instructions or contact our after-sales service. Did this help to solve your problem? YesNo Home
37
Check the power supply and connections. Did this help to solve your problem? YesNo Home
38
Either the screen or the machine is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
39
Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home
40
While carrying out a manual override, gently tap on the black screw next to the right hand pressure gauge; a solenoid valve may be stuck. Otherwise, the solenoid valve or the solenoid valve coil is defective. Contact our after-sales service. Did this help to solve your problem? YesNo Home
41
The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Possibly replace the regulator outlet coupling. Ref: 3.1.SD1 Did this help to solve your problem? YesNo Home
42
Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home
43
Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
44
Check the regulator setting by turning it anti-clockwise; if the pressure does not drop, the dosing module is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
45
Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
46
Defective solenoid valve coil. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
47
Leak in the pipe between the device and the tank. Repair it with a joint or replace it. Ref: 1.6Manchon Did this help to solve your problem? YesNo Home
48
Internal circuit blocked. Contact our after-sales service. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home
49
Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home
50
Enter a tank volume in hL and a dose in mL/L/month. Did this help to solve your problem? YesNo Home
51
Often it is an inputting error: enter the desired values again. Did this help to solve your problem? YesNo Home
52
The machine has lost its programme. Contact our after-sales service. Did this help to solve your problem? YesNo Home
53
Contact our after-sales service. Did this help to solve your problem? YesNo Home
54
Thank you for using Vivelys' online help. We hope that we have met your expectations, and are available to help if you need any further information. Please do not hesitate to contact us by email or telephone: 04.67.85.68.40 sav@vivelys.com Home
55
Do you need more help? Were you unable to solve the problem using the online troubleshooting guide? Please contact our technical support team by email or telephone: 04.67.85.68.40 sav@vivelys.com Home