1 Next Home Visio2 troubleshooting guide Start the slide show then click Next

2 VISIO2 after 2007 Home Symptoms of the problem (click on the relevant line) No control screen display. The control screen does not switch on or has frozen up. The oxygen consumption is too high. The control screen works but nothing happens, even if you override manually. The right hand pressure gauge is below 2.5 bars on all dosers. The right hand pressure gauge is above 2.5 bars on one of the dosers. The pressure on the right hand pressure gauge does not reach 2.5 bars on one of the dosers. The pressure on the left hand pressure gauge stays at 0 on one of the dosers. The pressure on the left hand pressure gauge is above 2 bars on one of the dosers. The Norton tube (noodle) on the outlet of one of the dosers disconnects constantly. The equipment will not run in automatic mode. The equipment does not work and the machine shows an error code. The right hand pressure gauge is below 2.5 bars on one of the dosers.

3 YesNo Home Is the equipment connected to a power supply?

4 YesNo Home Disconnect and reconnect the device to reinitialise it. Has the machine restarted?

5 YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)

6 YesNo Home Is the pressure on the right hand pressure gauge at 2.5 bars?

7 YesNo Home Is the equipment connected to the regulator properly? (sometimes the connection is poor due to oxygenation)

8 YesNo Home Is the equipment connected to the regulator properly? And is the pressure at 6 bars? (sometimes the connection is poor due to oxygenation)

9 Manually override in small increments by turning the regulator clockwise to 2.5 bars. Did this help to solve your problem? YesNo Home

10 YesNo Home Is the pressure on the left hand pressure gauge over 2.5 bars?

11 Completely unscrew the regulator, carry out a manual override in order to remove any impurities, then reset the pressure to 2.5 bars. If the problem persists, the doser module is defective. Contact our after- sales service. Did this help to solve your problem? YesNo Home

12 YesNo Home Carry out a manual override. Are the solenoid valves working?

13 YesNo Home Are the Norton tubes (noodles) on the doser outlet and the tank inlet disconnected?

14 If yesIf no Home Check the ceramic element; it may be broken or the pipe connecting to the diffuser may be disconnected.

15 YesNo Home Is the device's outlet coupling properly connected? (risk of a poor connection due to oxygenation).

16 YesNo Home Is the diffuser clogged?

17 YesNo Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas should flow. Does the left hand pressure gauge fall again immediately?

18 YesNo Home Is the diffuser clogged?

19 YesNo Home Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas should flow. Does the left hand pressure gauge fall again immediately?

20 YesNo Home Have the tank and dose volumes been programmed?

21 Yes Home Is the programmed dose compatible with the volume programmed? (max 120mL/L/month). No

22 If yesIf no Home Check that there are no leaks in the circuit (see attached procedure).

23 Defective machine or supply Contact our after-sales service. Did this help to solve your problem? YesNo Home

24 The equipment is ready for use. Did this help to solve your problem? YesNo Home

25 The O 2 supply circuit is blocked. Contact our after-sales service. Did this help to solve your problem? YesNo Home

26 The O 2 supply circuit is blocked. Contact our after-sales service. Did this help to solve your problem? YesNo Home

27 The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home

28 Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home

29 You can repair it with a stainless steel olive or replace the entire tube. ref: 215TA10MO Did this help to solve your problem ? YesNo Home

30 Replace the coupling. Doser outlet 2/4 (ref: 1.6SortieDoseur2/4). Did this help to solve your problem? YesNo Home

31 Clean the ceramic element (see attached procedure). Check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home

32 Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home

33 Clean the ceramic element (see attached procedure). Check that it is working properly in a bucket before putting it back in the tank. Did this help to solve your problem? YesNo Home

34 Pipe blocked - replace it. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home

35 The machine has lost its calibration value; you can check it on the "SETTINGS" screen. If this value is 0, you will need to reprogramme it. The machine is programmed in the "SETTINGS" menu: enter the code "8574" then enter the calibration value for each doser and reset the code to 0. Did this help to solve your problem ? YesNo Home 0.0 USER CODE SYSTEM CODE

36 Repair following the attached instructions or contact our after-sales service. Did this help to solve your problem? YesNo Home

37 Check the power supply and connections. Did this help to solve your problem? YesNo Home

38 Either the screen or the machine is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

39 Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home

40 While carrying out a manual override, gently tap on the black screw next to the right hand pressure gauge; a solenoid valve may be stuck. Otherwise, the solenoid valve or the solenoid valve coil is defective. Contact our after-sales service. Did this help to solve your problem? YesNo Home

41 The equipments outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation). Possibly replace the regulator outlet coupling. Ref: 3.1.SD1 Did this help to solve your problem? YesNo Home

42 Set the O 2 bottle outlet pressure to 6 bars. Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to operate. Did this help to solve your problem? YesNo Home

43 Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

44 Check the regulator setting by turning it anti-clockwise; if the pressure does not drop, the dosing module is defective. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

45 Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

46 Defective solenoid valve coil. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

47 Leak in the pipe between the device and the tank. Repair it with a joint or replace it. Ref: 1.6Manchon Did this help to solve your problem? YesNo Home

48 Internal circuit blocked. Contact our after-sales service. Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it. If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in it needing to be replaced. It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use. Did this help to solve your problem? YesNo Home

49 Defective dosing module. Contact our after-sales service for a replacement. Did this help to solve your problem? YesNo Home

50 Enter a tank volume in hL and a dose in mL/L/month. Did this help to solve your problem? YesNo Home

51 Often it is an inputting error: enter the desired values again. Did this help to solve your problem? YesNo Home

52 The machine has lost its programme. Contact our after-sales service. Did this help to solve your problem? YesNo Home

53 Contact our after-sales service. Did this help to solve your problem? YesNo Home