1 A simple modification to lower incidence of heart block with sutureless valve implantation 
Bobby Yanagawa, MD, PhD, Jorge Cruz, MD, Lyna Boisvert, RN, CV-ICU, BScN, Daniel Bonneau, MD  The Journal of Thoracic and Cardiovascular Surgery  Volume 152, Issue 2, Pages (August 2016) DOI: /j.jtcvs Copyright © 2016 The American Association for Thoracic Surgery Terms and Conditions

2 Figure 1 Top: Image of the Perceval sutureless valve (courtesy of LivaNova, Milan, Italy) and the width from the eyelet to the sealing collar edge (insert, between arrows). Middle: Transesophageal echocardiographic image showing the distance from the annulus to the lower edge of the sealing collar (between arrows): 4 mm in a patient without PPM (left) and 10 mm in a patient with PPM for heart block (right). Bottom: Distance from annulus to lower edge of sealing collar for all patients. The Journal of Thoracic and Cardiovascular Surgery  , DOI: ( /j.jtcvs ) Copyright © 2016 The American Association for Thoracic Surgery Terms and Conditions

3 Low valve implantation with respect to the annulus (between arrows) and need for pacemaker.