CENTRO DE METROLOGIA QUIMICA
2
CALCULATION OF UNCERTAINTIES FOR THE REPORT OF ANALYSES IN SIM. 8. P1
CALCULATION OF UNCERTAINTIES FOR THE REPORT OF ANALYSES IN SIM.8.P1.2 (CHLORINATED PESTICIDES IN HEXANE) 1
3
ANALYTICAL MEASUREMENT
Gas Chromatograph Perkin-Elmer Model 8000 series 2 with Electron Capture Detector and manual injector Column Phase, BP- 5 Col. Length, m Col. i.d., mm Col. film thickness, µm HEXACHLOROBENZENE gamma-HCH (g-BHC, LINDANE) HEPTACHLOR HEPTACHLOR EPOXIDE 2,4'-DDE CIS-CHLORDANE (ALPHA-CHLORDANE) TRANS-NONACHLOR DIELDRIN 4,4'-DDE 2,4'-DDD 2,4'-DDT 4,4'-DDT MIREX 1
4
ANALYTICAL MEASUREMENT
NIST-SRM 2261 was used for the preparation of calibration standards. The quantification method used was bracketing using the method of internal standard. Sample of SIM.QM-P1 was selected as control 1
5
ANALYTICAL MEASUREMENT
GENERAL SCHEME OF ANALYTICAL MEASUREMENT Gravimetric Preparation of standards, samples and controls Gravimetric Preparation of dilution Verification of the standards through the calibration curve Determination by bracketing using the method of internal standard Statistical evaluation of the results and uncertainty calculation 1
6
GRAVIMETRIC PREPARATION OF STANDARDS
7
BRACKETING SCHEME 120 58 40 20 8 4,4 ‘ -DDE
ng/g 120 58 40 20 8 4,4 ‘ -DDE HCB, g-HCH, HEPTACLOR EPOXI, MIREX, 2’4-DDT, 4’4-DDT DIELDRIN HEPTACLOR (s/d), cis-CLORDANO, trans-NONACLOR, 2’4-DDD, 2’4-DDE 1
8
THE UNCERTAINTY ESTIMATION PROCESS
9
BRACKETING AND INTERNAL STANDARD
(AH/AIS) (AS/AIS) (AL/AIS) CS/CIS CH/CIS CL/CIS Sample concentration is calculated trough the equation: 1
10
IDENTYFYING UNCERTAINTY SOURCES
Csample
11
QUANTIFYING UNCERTAINTY SOURCES
Csample
12
UNCERTAINTY OF THE STANDARDS
13
UNCERTAINTY OF THE STANDARDS
15
GRAVIMETRIC PREPARATION OF STANDARDS
Ampoules SMR 2261 sol. INT 1 sol. INT 3 Standard Brack low Brack high IS – sol. INT IS solid 0.5% 0.6% 0.6% 0.6% 0.6%
16
IDENTYFYING UNCERTAINTY SOURCES
Csample
17
IDENTYFYING UNCERTAINTY SOURCES
Csample
18
PURITY OF THE INTERNAL STANDARD
The purity of IS is quoted by the supplier as 99.0 % . The quoted uncertainty is taken as a rectangular distribution
19
QUANTIFYING UNCERTAINTY SOURCES
20
UNCERTAINTY STANDARD OF THE SIGNAL
The uncertainties associated with the variability of the area were estimated by making replicate measurements of samples and bracketing standards. These uncertainties are estimated through the following equation:
21
UNCERTAINTY STANDARD OF THE SIGNAL
22
QUANTIFYING UNCERTAINTY SOURCES
23
CALCULATING THE CONCENTRATION
25
UNCERTAINTY CONTRIBUTIONS
26
CONCENTRATION AND UNCERTAINTY VALUES FOR THE SAMPLE SIM.8.P1.2