Ch6 Measurement of Ionizing Radiation

Similar presentations


Presentation on theme: "Ch6 Measurement of Ionizing Radiation"— Presentation transcript:

1 Ch6 Measurement of Ionizing Radiation

2 Introduction Chemical and Biologic effect Skin erythema dose (SED)
AD 1928 , ICRU , Roentgen Exposure(X)

3

4

5 Free-Air ionization chamber Correction for
(a) air attenuation (b) recombination of ions (c) the effects of temperature, pressure, and humidity (d) scattered photons

6 Limitations : High energy photons
Non-unique electric field and greater ion recombination

7 Thimble chambers The inner cavity is equal to the maximum range of electrons

8 Effective of Atomic Number

9 Chamber Calibration (A) too few electrons are generated in the wall
(B) reduced because if increased attenuation of the beam in the wall

10 Practical Thimble Chambers
Condenser Chambers

11 Stem Effect

12 Farmer Chamber Condenser chamber--- lower energy beams (<=2MV)
The design of stem and excessive stem leakage create dosimetric problems

13

14

15 String Electrometer

16 Other Electrometer

17 Special chamber- Extrapolation Chamber

18 Ion Collection Incomplete ion collection at low voltages

19 Collection Efficiency
Two voltage testing techniques Ion recombination correction

20 Chamber Polarity effects
Compton current Extracameral current Environmental conditions Exposure calibration factor(Nc):NIST Ctp:Temperature and pressure Cs:Recombination Cst:Stem leakage correction