Reference Membrane Location
2
Reference Membrane
3
Span Check Perform 8 minute count Compare mass value with ABS value
ABS must be within 5% or BAM will log CAL error (D error code)
4
Calibration Test Screen
CALIBRATION MODE REF MBRN: < COUNT (Io): 634000 COUNT (I): 556234 CAL MASS M: 0.801 mg/cm2 START STOP EXIT
5
Hands On Workshop – Reference Membrane
Remove Reference Membrane Inspect Reference Membrane Reinstall Reference Membrane Inspect mechanical movement of Reference Membrane Perform Reference Membrane Count/Mass Test in TEST/CALIBRATE Screen Compare Mass value with ABS value