1 Hussam Ali et al. JACEP 2016;2:118-119
RBBB and Pacemaker Malfunction (A) Electrocardiogram recording showing inappropriate ventricular pacing (solid arrows) following the appearance of right bundle branch block (RBBB) (dotted arrow). (B and C) Bipolar electrograms recorded by the ventricular lead before and after RBBB, respectively. Solid arrows indicate inappropriate pacing, and asterisks indicate premature ventricular beats. (D and E) Chest radiographs with presumed propagation patterns (dotted arrows) before and after RBBB (double line), respectively. Hussam Ali et al. JACEP 2016;2: American College of Cardiology Foundation