1
Air-Water (n=1.3)-Air film (soap film) Normal incidence
Can we make a device with greater contrast in transmission? T Dd/l
2
Air-Diamond (n=2.4)-Air film
T F = 0.88 Dd/l
3
Air- Kryptonite? (n=10) -Air
F = 18 Dd/l Interference ideas for why increasing r makes these peaks so sharp?
4
Multiple reflection view of T at low angle
As long as |x| < 1, this will converge But for |x| 1, the dependence on phase is very fast!
5
If |r|,|r’| approach 1, many reflections contribute to the interference for R and T. Many phasors of nearly equal size give much more contrast in magnitude and much sharper features than just a few of very different size. Very important concept in many contexts of interference
6
Multiple reflection view of T
Suppose R = rr = The relative strength of the first three transmitted fields are 1, 0.1, 0.01… etc . If added in phase (0 shift) you get ____ If added out of phase (p shift) , you get ____ Max/min fields______ Max/min T’s _______
7
Multiple reflection view of T
Suppose R = rr = The relative strengths of the transmitted fields are 1, 0.75, …. etc . Now we must add many more fields to get the transmission.
8
Multiple reflection view of T
Sketch with phasors the addition of the fields 1, 0.75, …. etc in phase This adds up to something close to 2 4 6 8 10
9
Multiple reflection view of T
Sketch with phasors the addition of the fields 1, 0.75, …. Out of phase (p shift) This adds up to something close to 0.1 0.2 0.4 0.6 1 Max/min fields______ Max/min T’s _______
10
Multiple reflection view of T
Sketch with phasors the addition of the fields 1, 0.75, …. at p/4 shift This adds up to something close to 0.2 0.4 0.6 1 2 Do the same for 1, 0.1, 0.01 In a relative sense, which is most sensitive to phase?
11
General feature of wave interference: if you want sharp features you have to add lots of waves with varying phase. (e.g. Fourier)
12
If I want to make an instrument that distinguishes wavelengths, I can get sharp transmission by increasing n to increasing R. But it’s hard to obtain and work with kryptonite.
13
Fabry-Perot Etalon/Interferometer
High R due to large k in in metal coating.
14
Fabry-Perot Etalon/interferometer for low R:
Fabry-Perot Etalon/interferometer for high R, same d Red: l = 1.00 Blue: l = 1.01 Is the device most sensitive to l changes for small or large d?
15
Free spectral range: At a given d, how much can l change before it confuses us by “lapping” its original peak near d?
16
Given a choice, is it better to increase d or increase R to make as small as possible?
a) increase d b) increase R
17
Multiple reflection view of T
18
Solar corona taken with the LASCO C1 coronagraph which employed a tunable Fabry-Pérot interferometer near an iron atom emission line at nm. Color coded image of the doppler shift of the line, after subtracting solar rotation.
19
What you see on a screen for any interferometer
There is always some angular divergence
20
Two close sodium emission lines are visible
Experiment on video