1 Volume 7, Issue 8, Pages 1048-1055 (August 2010)
Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures  Linda M. Kallinen, BS, FHRS, Robert G. Hauser, MD, FHRS, Chuen Tang, MD, Daniel P. Melby, MD, Adrian K. Almquist, MD, William T. Katsiyiannis, MD, Charles C. Gornick, MD  Heart Rhythm  Volume 7, Issue 8, Pages (August 2010) DOI: /j.hrthm Copyright © 2010 Heart Rhythm Society Terms and Conditions

2 Figure 1 Top: Gradual impedance rise culminating in impedance spikes in a patient (no. 23; Table 1) without lead integrity alert LIA). Middle: Stable impedance over 14 months followed by accelerating impedance that triggered LIA (no. 12; Table 2). Bottom: Gradual rise in impedance in a patient who had received appropriate implantable cardioverter-defibrillator therapy (no. 2; Table 2). RV = right ventricular. Heart Rhythm 2010 7, DOI: ( /j.hrthm ) Copyright © 2010 Heart Rhythm Society Terms and Conditions