Bone Density 101 Ben Williams DEXA Solutions, LLC
BS Biomedical Engineering (SLU) 17 years in the DXA industry
2
DXA – Dual-Energy X-Ray Absorptiometry
Most common method for determining bone density Why not an exact science? BMD = BMC (g)/A (cm^2) Error Variations between OEMs Differences in technologists Service performed incorrectly T-Score and Z-Score -1 and above = normal Between -1 and -2.5 = Osteopenic -2.5 and lower = Osteoporosis Z-Score of 0 is average
3
DXA – Dual-Energy X-Ray Absorptiometry (continued)
Common measurement sites (reimbursed) AP Spine Femur (Dual Femur) Forearm LVA (lateral vertebral assessment) Reimbursement varies Additional Applications (currently no reimbursement) Total Body and body composition CoreScan Pediatric Orthopedic
4
Why a Needed Technology?
Worldwide, osteoporosis causes more than 8.9 million fractures annually, resulting in an osteoporotic fracture every 3 seconds Osteoporosis is estimated to affect 200 million women 1 in 3 women over 50 will experience osteoporotic fractures, as will 1 in 5 men $14 Billion/year cost in US only
5
OEMs and Models of Densitometers
GE Lunar iDXA Prodigy DPX (NT, Duo, Bravo) DPX (IQ, alpha, L) Peripherals (Achilles) Hologic Discovery Series Horizon QDR Series Peripherals (Sahara) Norland XR Models
6
System Overview – Lunar Basic Function and Setup
Lunar systems are controlled by DOS, WIN95/98/XP/7 workstation using the Encore Software (does have a service mode for troubleshooting and calibration – different passwords for different versions). PC communicates to SBC Board via RS422/232 Serial card SBC (controller for all modern models) gives commands (motion, x-rays, etc…) Systems use 110V AC (20A dedicated outlet recommended by OEM) Isolation transformer used to limit leakage currents UPS used for isolation transformer and Scanner power Printer for reports (unless configured to send electronically)
7
System Overview – Lunar 3 Sub Systems
X-Ray Generation Tube Power Supplies X-Ray Detection CZT Detector Detector Electronics Scanner Control Motion Control Error Detection Communications with Host PC / Detector
8
System Overview – Lunar Schematic
9
System Overview – Lunar Common Failures
Tube Arc Typically the tube, but could be caused by HVPS HVPS Enable and Disable failures Broken Cable Shutter Failures PC Failures Deteriorated Detector (NT, Duo, Bravo, IQ)
10
System Overview – Lunar Troubleshooting
Error Log Tools – Error Log – Print all errors Arcing Look for oil leaking from tube’s expansion chamber Older models may need regreased O-ring inside collimator deteriorates Determine sub system (x-ray generation, detection or scanner control) Verify voltages on DC Supply (+/- 12V and 5V) Places to check voltage: DC Supply, DIB Board; Bulkhead connection
12
System Overview – Hologic Troubleshooting
13
System Overview – Hologic Troubleshooting
14
System Overview – Lunar Connectivity
Printed Reports Generally using a color printer PACS Configure DICOM and Worklist EMR (HL7 Work-Around) Install Adobe Writer Install, or connect to, EMR software Create reports as .pdf Upload to EMR
15
DXA Market Today Reimbursement down 70% in some areas
Many PC facilities not maintaining densitometers Shift back toward traditional market Used and refurbished markets growing
16
DXA Market Tomorrow DXA offered primarily in traditional market facilities Increase in availability overseas Huge push in “preventative care” Body Composition Type to enter a caption.
17
If your having this kind of a day let us help.
18
Questions??? Ben Williams President DEXA Solutions, LLC