Arm-to-arm variation when evaluating neuromuscular block: an analysis of the precision and the bias and agreement between arms when using mechanomyography or acceleromyography C Claudius, L.T. Skovgaard, J Viby-Mogensen British Journal of Anaesthesia Volume 105, Issue 3, Pages (September 2010) DOI: /bja/aeq162 Copyright © 2010 The Author(s) Terms and Conditions
2
Fig 1 Bland–Altman scatter plot of the relative bias in % against the mean times to TOF 0.90 at the two arms using mechanomyography. The bias increases as the mean time to TOF 0.90 increases. Therefore, all data are logarithmically transformed and the bias is presented as the relative bias (%). Data are also presented in Table 1. Four patients were antagonized before TOF 0.90, and in one patient, the monitoring was stopped at TOF 0.88 and 0.90 at dominant and non-dominant, respectively. Therefore, only 15 data sets are presented. British Journal of Anaesthesia , DOI: ( /bja/aeq162) Copyright © 2010 The Author(s) Terms and Conditions