Intravascular fracture of the Impella device during removal

Similar presentations


Presentation on theme: "Intravascular fracture of the Impella device during removal"— Presentation transcript:

1 Intravascular fracture of the Impella device during removal
Simon A. Castro, MD, Jackson J. Liang, DO, Erica S. Zado, PA-C, FHRS, Amit K. Mehrotra, MD, Francis E. Marchlinski, MD, FHRS, Pasquale Santangeli, MD, PhD  HeartRhythm Case Reports  Volume 3, Issue 12, Pages (December 2017) DOI: /j.hrcr Copyright © 2017 Heart Rhythm Society Terms and Conditions

2 Figure 1 A: Arteriogram of left common femoral artery showing occlusion. B: Catheter piece initially removed from body. Note that the catheter had sheared off at the pump motor. C: Fluoroscopy during surgical removal of the retained piece of catheter (arrow points to retained fragment). D: Retained fragment removed at surgery: distal end of catheter including motor pump and blood inlet area. HeartRhythm Case Reports 2017 3, DOI: ( /j.hrcr ) Copyright © 2017 Heart Rhythm Society Terms and Conditions


Download ppt "Intravascular fracture of the Impella device during removal"