Optimisation in proton scanning beams
Optimization in proton scanning beams Optimisation in proton scanning beams First results… Copyright (c) Varian Medical Systems Evangelos Matsinos, May 2002
2
Basic steps Calculation of the dose-deposition coefficients (ddc’s)
Optimisation of the spot weights
4
Spots and beamlets Beam beamlets or pencil beams (defined by the resolution of the calculation grid) The dose from each beamlet is evaluated (at the vertices of the calculation grid) The spot dose is calculated (as the sum of the dose contributions of the corresponding beamlets, weighted for the position of each beamlet within the spot)
5
Sx and Sy depend on z spot depends on the density and on z beamlet
6
Optimisation: a closer look
Desired dose at point i: pi Dose delivered at point i: di = aij xj (sum over all sources j) Objective function: Fobj = (di - pi)2 (sum over target points) + contribution due to the violation of dose-limit constraints (for targets and organs) + contribution due to the violation of dose-volume constraints (for organs)
7
Optimisation methods Conjugate Gradient (CG) Simulated Annealing (SA)*
‘Simultaneous’ optimisation (PSI) Generalised Sampled Pattern Matching (GSPM)* (* = under development)
8
Strategy in the optimisation
Pre-optimisation Reasonable initial ‘guess’ for the weights Convergence two consecutive iterations yield improvement below 5% Main optimisation Full implementation of a method yield improvement below 0.1%
9
Toy example A phantom has been created with three important structures: one target and two organs; some inhomogeneity has been introduced (an additional structure simulating the presence of a bone) Pixel size: 2.5mm Spot advance in y (scanning direction): 2.5mm Spot advance in x: 5mm Cut-off for dose contributions: 3 standard deviations
10
Distal Organ: 2,166 points, 48.34 cm3
Target: 2,412 points, cm3 Distal Organ: 2,166 points, cm3 Proximal Organ: 683 points, cm3 Number of points: 5,261
11
Number of parameters: 4,798
12
ProtonHelios
13
Dose-Volume Histograms
Prescription dose: 50 Gy ( 2%) Organ constraints: 25 Gy in 10% of the distal organ;15 Gy in the proximal organ
14
Dose distribution (exclusive fit to the target)
15
Dose distribution (fit to all structures)
16
Comparison of a few numbers
Method Minimal Fobj Relative time Target dose (Gy) Maximal weight CG 36,902 3.87 49.6 2.3 16.03 SA 35,621 4.92 17.15 PSI 37,440 1.00 49.7 2.3 51.97 GSPM --- 1.33 49.9 1.9 6.74
17
Weight distribution PSI method
18
Weight distribution SA method
19
A head tumour
20
Conclusions As far as the dose distribution is concerned, three optimisation methods (CG, SA, and PSI) yield results which seem to be in good agreement. Very similar dose distributions may be obtained on the basis of very different weight distributions. The use of raw (unfiltered) weights does not seem to create cold/hot spots within the irradiated volume. It remains to be seen whether, in some occasions, filtering will be called for.
21
Under consideration… Other forms of the objective function to be tried? Strategy in the optimisation: an improvement of about 25% was found in the execution time in case that the target dose is firstly optimised (with vanishing dose everywhere else) Other optimisation methods to be tried?