Download presentation
Presentation is loading. Please wait.
Published byHadian Sutedja Modified over 5 years ago
1
C100 Operation Enhancing gradient reach for the
12 GeV physics program via fault management
2
We are still learning and pressing forward
Overview From commissioning to the present, C100 gradient maximization while maintaining an acceptable trip rate has been… challenging. Many avenues for enhancing cryomodule performance have been discovered, explored and acted upon by many contributors from multiple divisions. Intra-divisional collaboration has been key in investigating and resolving issues And - while some mysteries remain in fully mastering C100 performance to realize true and sustainable 12 GeV operation – To date… The five C100 modules in each linac provide about 40% of the energy gain required Highest gradients to date for cryomodules in a CW machine Beam quality has been acceptable and most 12 GeV era milestones have been attained We are still learning and pressing forward
3
Brief history of issues
Heaters Went from 4 to 8 Individual cavity heater control installed in injector V2 chassis now being tested for remaining zones Tuner performance Bolts installed in non-tensioned tuners Went from EPICS to fiber input, faster updates – tighter tuning Tightened dead bands Waveforms give real-time detune information SEL to GDR transition Automated turn on and recovery Implemented 27 degree shift for POFF Service building temperature excursions Caused the 56 MHz clock to lose lock Crosstalk Re-terminated many heliax cables Still present as regulation degradation (across cavities and zones)
4
Brief history of issues, cont’d.
Microphonics Stiffened tuner lever arms Construction obstruction Tom Powers, Ken Baggett, George Biallis and Bob Legg tackled this one – wiggly waveguide Solution most apparent in 0L04 More hardening in the future Helium pressure standardization Helium vessel design for heat dissipation = 2.07K (0.037 Atm) Had to research spec C100 placement = higher return pressure than at “T” Further testing (Freyberger, et al) suggests higher return pressure may be sufficient, which puts less stress on CHL
5
Present concerns Particulate contamination Field Emission
Dark current generation Radiation damage Quench at lower gradient than commissioning limits “Anomalous” Quench – (field depletion due to non-classic quench)
6
Types of faults “Hard” RF is turned off due to fault detection
SEL quench Arc Vacuum Quench Others… “Soft” Cavity transitioned to SEL mode to reduce recovery time GLDE GDCL DETA PLDE Every fault (hard or soft) pulls an FSD
7
GDCL “soft faults”
8
Quench Faults
9
So why are the C100 cavities quenching so often?
Focus on quench As we know, there are multiple fault flavors to contend with. There may be some interaction between detuning and the “soft” faults, but “quench” is concerning since the cavities were commissioned with gradient constraints below the field potential where a quench was induced – and this was historically reliable as long as a new field emitter didn’t turn on… So why are the C100 cavities quenching so often?
10
Quench Algorithm E field collapses from cavity faster than the Qext decay rate Derived from probe signal Calculate gradient reduction in counts for 2 MV/m Find time of gradient loss in msec Monitor gradient loss slope based on these values
11
Classic quench event 600 usec decay
12
Different “quench” event
50 usec decay??
13
Multi – cavity response to quench event
14
Quench rate mitigation – 1L23
15
Cavity field depletion aka “quench” - Questions
Are there multiple mechanisms (besides classic “quench”) that induces cavity field depletion at a rate faster than the normal decay time of the cavity? (Arcs as Tom alludes to?) End group heating? (1L23-8?) Are the quenches being detected “real” or is another mechanism at play? Is the algorithm for quench correctly applied? Regardless of the cause for quench detection – it still counts as a trip, and the only defense we have is to reduce operating gradient….
16
Closing comments Quench faults dominate the trip rate of the C100 cavities. There are other areas of concern that requires further investigation, however, if the quenches, arcs, or whatever mechanisms that cause rapid field depletion can be defined (or perhaps mitigated), that information could be used to develop a strategy towards maximizing gradients and minimizing trip rates in the C100 zones. Thanks to all the people who provided information and assistance to help us understand unfamiliar concepts.
17
End
18
GDCL Soft Fault
19
GLDE Soft Fault
20
Day to day operation Goal – provide desired energy at the lowest trip rate For C100 zones - How do we get there?
21
Now for something completely different
Original Display
22
FSD Count – Is this acceptable?
24 hour period of FSDs from C100 cavities
23
Different, cont’d.
24
Still different
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.