Simulated Distillation D2887, ISO3924, IP406

Similar presentations


Presentation on theme: "Simulated Distillation D2887, ISO3924, IP406"— Presentation transcript:

1 Simulated Distillation D2887, ISO3924, IP406
Boiling Point Determination for Petroleum Fractions

2 SIMDIST D2887, ISO3925, IP406 Product Value Proposition
Competitive Positioning Sales Tools How to Order Demonstrated Analysis

3 Product – SIMDIST D2887

4 SIMDIST Purpose Complete TBP analysis
Establish Boiling Point Curve for petroleum fraction Product value ($$) Product compliance – measurement is often mandatory! 90%OFF is essential to ensure product is OK. Reference Method: D86 (very old, 70+yrs)

5 SIMDIST Principle Gas Chromatography key attributes
Boiling Point Separation FID mass detector Retention time vs Boiling Point. Run Calibration Curve: n-parrafins C5-C44 (Carbon # vs BP) Run Blank: establish clean baseline for subtraction. Run Reference Sample: signal vs RT  %mass vs BP Run Unknown Samples, same as Reference. Report %Mass vs BP, fractions (i.e. 90% OFF)

6 SIMDIST Principle Peak ID Compound Name BP (°C) 1 n-C5 36 2 n-C6 69 3
98 4 n-C8 126 5 n-C9 151 6 n-C10 174 7 n-C11 196 8 n-C12 216 9 n-C14 254 10 n-C15 271 11 n-C16 287 12 n-C17 302 13 n-C18 316 14 n-C20 344 15 n-C24 391 16 n-C28 431 17 n-C32 466 18 n-C36 496 19 n-C40 522 20 n-C44 545

7 SIMDIST Principle

8 <1% (not essential though!)
Analyzer Scope D2887(A/B), ISO3925, IP406 Component Anaytical Range Precision (%) Range / Scope Up to FBP 538° Refer to Method Min Boiling Range 55.5°C Max FBP (99.5% point) 538° Sample Types Can be analyzed? Diesel Yes Jet fuel Biodiesel (B5, B10, B20) Only D2887-B Accelerated Gasoline or Blend Streams Not suitable – use D7096! Injection Repeatability <1% (not essential though!) Analysis time 35 mins / 8 mins

9 Configuration A Scion SIMDIST Analyzer includes:
436-GC or 456-GC including Liquid CO2 cryogenic oven cooling. COC injector w. EFC and air cooling High Temp FID detector with DEFC 8400 or 8410 AutoSampler (optional Cooling) PC with CompassCDS workstation software & SIMDIS Plug-in 10m Column All included packages, w/o Compass optional . Tested & Documented

10 SIMDIS – Value Proposition

11 Value Proposition SIMDIST D2887, ISO3925, IP406
Dedicated Solution Designed, Built, Tested & Documented to method Platform Options; 436 Smallest Footprint 456 Flexible Platform - Allows dual method (e.g. D D7096) Accuracy - Trusted data, Maximum Product Value! Proven Results Software for automation, calculus, reporting Better ROI / Lower Cost of Ownership Fast Cycle Times – especially with D2887 Accelerated! Low maintenance Compliancy to Method D2887 Procedures A+B (both can be run using same column)

12 SIMDIST - Options

13 D2887 or D2887 Accelerated? D2887 has 2 procedures (since 2018):
Procedure A: Classic Method. 10M*0.53*2.65µm Procedure B: Accelerated – faster programming of GC oven, 10M*0.53*0.88µm Differences A vs B Runtime: A is 35 mins, B is about 8 mins Scope: B will also allow for Biodiesels up to B20. Procedure A is not used for Biodiesels Bias: Small bias between A and B. Method specifies and one may correct. Precision: Comparable Full details in Method Both D2887 available in Scion. In many cases, Accelerated is perfectly fine for customer (and therefore recommended)! Switching between regular and Accelerated always possible Change column, method

14 CP-8400 and CP-8410 AutoSamplers
Key Features CP-8400: 100x2mL CP-8410: 10x2mL / 6x5mL / 5x10mL Two Injector Access Dual/Duplicate Injections Wide range of Syringes Three Solvent Wash Upgrade to Large Rinse Vial Waste Drain Tray Heating/Cooling (Option) Flow Cell (Optional)

15 Software Options CompassCDS Eclipse plug in Dedicated tool
Clear Interface Easy to use Multiple dedicated panels Easy Key Graphical Functions Icon tool strip, Zooming/Selecting/Dragging

16 Software Options CompassCDS Eclipse plug in
Standardized Methods (editable) Interval report in °C or °F (editable) Alkane/Carbon # profiling

17 Software Options CompassCDS Eclipse plug in
Volume% correlations for D2887 to D86, STP577, D1160 (correlation models editable) D93 Flashpoint MOV D6417 Noack Evaporation Loss - DIN Validation Reports (system check)

18 Other Options Additional channels Another SIMDIST channel (e.g. D3710)
for dual SIMDIS Other analytical channels e.g. a DHA channel… Be ware of Temperature limitations of respective channels!!

19 Compliancy D2887 Other SIMDIST Methods
Standard Test Method for Boiling Range Distribution of Petroleum Fractions by Gas Chromatography. Other SIMDIST Methods SIMDIST Mode Kits are available to switch method quickly to perform any other SIMDIS method.

20 Competitive Positioning

21 SIMDIST methods Overview
* method withdrawn. Now included in D7096 **method withdrawn. Method Sample Type / Scope Max Carbon # IBP (°C / F) FBP (°C / F) Col. Length (m) Analysis time (min) D3710* Gasoline, Naphtha C15 n/a 260 500 D7096 C16 280 536 D2887 Jetfuel, Diesel C44 538 1000 10 35 D2887 Accelerated ISO3924 IP406 D5307* Crude D5442 Waxes D7213 Lubes C60 100 212 615 1138 D7398 Biodiesel Blends D6352 C90 174 345 700 1292 D7169 Residuals, Crude Oil C100 720 1328 5 D7500 C110 735 1355 IP480 Lube Oil Base (100% eluting) C120 750 1382 EN DIN 51435 IP507 EN IP545 Crude Oil EN *obsolete/withdrawn

22 SIMDIST - Things to remember!
Obsolete methods D3710 (7096) and D5307 withdrawn. D2887 procedure A vs B (accelerated) Faster Small bias (due to Temp programming)! Compliancy Methods are standardized for a particular product type, IBP/FBP! Biodiesel blends could impair results (also for D86) Local Governing Local body will mandate physical or GC or both allowed.

23 Other Solutions/Methods
Physical Distillation D86 Local regulations always will dictate what method to use!! 8612 GC method DHA style method, modeling to D86 (Fugacity model). D86 Gr 1/2 only Gasoline + Blends, Naphtha. Not suitable for Jet/Bio/Diesel) Correlation to D86 Fast method (8 mins) Non-standardized method!

24 D2887 vs D86 Physical Distillation
SIMDIST Automatable Safe Easy to learn Very Simple hardware Better precision of data (typically 2-3x better ). Wider BP range / application range TBP correlation D86 Standard method – golden reference True TBP. Safety precautions required. Lower productivity, no automation Manual operation, attended Cost effective hardware higher afterward labor… Requires vacuum for Higher Temps cannot do crudes.

25 Competitive Landscape
GC 436 smallest footprint in market 456 for dual channel / dual method SIMDIS (or even triple…) All (D)EFC – Retention Time Specification comparable to other brands! Sampler Tray Heating/Cooling (for very volatile/viscous samples!) Dual/Duplicate capability Sampler really is required for SIMDIS Compass w Eclipse Easiest to use & learn Dedicated Plugins for SIMDIS reporting 21CFR11. Scalable from single GC to full Client Server

26 Sales Tools

27 Sales Tools PPT App Notes Poster Board Demonstrated Analysis
Customer Value How to order Competitive App Notes Poster Board Demonstrated Analysis Hydrocarbons Ebook

28 How to Order

29 How to Order – 436-GC P/N DESCRIPTION 4369940946
SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 230V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 230V ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes PC, CompassCDS with license and Eclipse SIMDIS Plug-in software.

30 How to Order – 436-GC w/o PC & Compass
P/N DESCRIPTION SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 230V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 230V ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes SIMDIS plug in software CD. Excludes PC and CompassCDS.

31 How to Order – 456-GC P/N DESCRIPTION 4569940946
SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 230V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 230V ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes PC, CompassCDS with license and Eclipse SIMDIS Plug-in software.

32 How to Order – 456-GC w/o PC & Compass
P/N DESCRIPTION SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 120V SimDist Analyzer with 8410 for ASTM D2887, ISO 3924 & IP 406, 230V SimDist Analyzer with 8400 for ASTM D2887, ISO 3924 & IP 406, 230V ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes SIMDIS plug in software CD. Excludes PC and CompassCDS.

33 How to Order – 436-GC D2887 ACCELERATED 0N 436 - NEW P/N DESCRIPTION
SimDist Analyzer with 8410 for Fast D2887 Application, 120V SimDist Analyzer with 8400 for Fast D2887 Application, 120V SimDist Analyzer with 8410 for Fast D2887 Application, 230V SimDist Analyzer with 8400 for Fast D2887 Application, 230V ASTM D2887 Accelerated is Procedure B in D2887 with runtime <10 minutes; it is very useful for refineries to speed up the analysis time. The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. Includes PC, CompassCDS with license and Eclipse SIMDIS Plug-in software.

34 How to Order – 436-GC w/o PC & Compass
D2887 ACCELERATED 0N NEW P/N DESCRIPTION SimDist Analyzer with 8410 for Fast 2887 Application, 120V SimDist Analyzer with 8400 for Fast 2887 Application, 120V SimDist Analyzer with 8410 for Fast 2887 Application, 230V SimDist Analyzer with 8400 for Fast 2887 Application, 230V ASTM D2887 Accelerated is Procedure B in D2887 with runtime <10 minutes; it is very useful for refineries to speed up the analysis time. The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. Includes SIMDIS plug in software CD. Excludes PC and CompassCDS.

35 How to Order – 456-GC D2887 ACCELERATED P/N DESCRIPTION 4569940952
SimDist Analyzer with 8410 for Fast 2887 Application, 120V SimDist Analyzer with 8400 for Fast 2887 Application, 120V SimDist Analyzer with 8410 for Fast 2887 Application, 230V SimDist Analyzer with 8400 for Fast 2887 Application, 230V ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes PC, CompassCDS with license and Eclipse SIMDIS Plug-in software.

36 How to Order – 456-GC w/o PC & Compass
D2887 ACCELERATED P/N DESCRIPTION SimDist Analyzer with 8410 for Fast 2887 Application, 120V. SimDist Analyzer with 8400 for Fast 2887 Application, 120V. SimDist Analyzer with 8410 for Fast 2887 Application, 230V. SimDist Analyzer with 8400 for Fast 2887 Application, 230V. ASTM D The determination of the boiling range distribution of petroleum products. It is applicable to petroleum products and fractions having a final boiling point of 538°C (1000°F) i.e. carbon number C44 or lower at atmospheric pressure as measured by this test method. The method is limited to samples having a boiling range greater than 55°C (100°F). The method is not for gasolines. Use D-7096 for gasolines. Includes SIMDIS plug in software CD. Excludes PC and CompassCDS.

37 How to Order – Sampling Options
P/N DESCRIPTION CP739371 8400 AutoSampler for 45x-GC/43x-GC Provides maximum throughput with a standard 100 sample carousel for 2ml vials; automatic access of two injectors with a single tower in dual and duplicate injection mode; headspace sampling or large volume injection with 100ul syringe; variable sampling depths for headspace and micro extracts; internal standards or matrix spikes can be automatically added to any injection; wash station for 3 solvents-programmable in any combination; optional sample heating and cooling 10-40°C; a menu of pre-programmed standard injection parameters minimize method development time. The sampler accepts syringe sizes from 1 up to 250µL. Includes: 100 vial (2mL) capacity carrousel, 10ul syringe, 100 x 2ml vials with caps and septa, wash station for 3 solvents, 4 x solvent wash bottles and caps and septa, one waste vial. CP739651 8410 Auto Injector for 45x-GC/43x-GC GC Provides flexibility with a fixed sample tray supporting multiple vial sizes; automatic access of two injectors with a single tower in dual and duplicate injection mode; headspace sampling or large volume injection with 100ul syringe; variable sampling depths for headspace and micro extracts; internal standards or matrix spikes can be automatically added to any injection; wash station for 3 solvents-programmable in any combination; a menu of pre-programmed standard injection parameters minimize method development time. Includes: Ten-2mL, six- 5mL, and five-10mL sample vial carrousel with standard 10uL syringe; wash station for 3 solvents, 4 x solvent wash bottles and caps and septa, one waste vial. Note: in all std DHA packages, 8400 or 8410 is included.

38 How to Order – Sampling Options
P/N DESCRIPTION CP741161 84x0 Large Solvent Vial Additional 120 mL stainless steel vial. CP739379 8400 Carousel Cooling/Heating Plate ( °C) Requires circulating bath. CP740608 8410 Carousel Cooling/Heating Plate ( °C) DY Re-Circulating Bath, 120V. DY Re-Circulating Bath, 230V, CE mark.

39 How to Order – Mode Kits P/N DESCRIPTION CP741070 D7096 Application
D7213 / DIN Application CP741073 D2887 accelerated Application CP741074 High Temp / ASTM D6352 / D7169/ D7500 / IP 480 Type A/B/C Application ASTM D 7096 Application The SimDist Mode Kits include all parts to change from one standard configuration to another one. Each kit contains: column, syringe, liners, septa, samples, flame tip & ferrules.

40 How to Order - Software P/N DESCRIPTION DP-PC-XX-SI-10
SIMDIST Software for CompassCDS, 1 Instrument Requires CompassCDS or CompassCDS demo software version 2.0 or higher. Order CompassCDS separately. Not compatible with Galaxie. RECOMMENDED: On-Site Training (1 day) in addition to Workstation training, to be offered separately. DP-PC-XX-SI-11 SIMDIST additional license Note: in all std SIMDIS packages, Plugin is included with License.

41 How to Order - Consumables
P/N DESCRIPTION SC37787 SCION-Simdist Inert Steel 10m x 0.53mm x 0.88Um (D2887 accelerated) SC37788 SCION-Simdist Inert Steel 10m x 0.53mm x 2.65Um (D2887 regular) CP741030 SimDist D-2887 Reference Standard (6 x 1ml, neat) CP741031 SimDist D-2887 Calibration Standard (1 x 1ml, diluted) SG092034 0.8mm ID restriction INSERT SPI, 0 .53mm ID, 5/PK CR298777 Septa BTO, 11.5 mm CenterGuide™, 50/PK CP743117 Capillary column nut stainless steel for 1093 injector CR211108 FERRULE, Graphite, 1/16" to 0.8mm ID, 10/PK, for 0.53 mm ID Columns HM202999 Syringe, 5 µL, Hamilton CP741148 Particle filter for LCO2 option

42 How to Order - Spares P/N DESCRIPTION CP745238
Fuse F-15A/250V, 5x20mm, Glass (24V DC,3AG) BR746381 Fuse transformer 120V F-10A 250V 6.2x32mm BR746382 Fuse transformer 230V F-8A 250V 6.2x32mm Fuse F-15A / 24V (3AG) CP745239 Fuse T- 10A/250V, 5x20mm, Glass HBC (Heaters) CP745240 Fuse T- 15A/250V, 5x20mm, Glass (Column Heater) BR746360 Chassis Fan Column Oven Motor, 120V Column Oven Motor, 230V Vent Door Stepping Motor Column Oven Heater, 120V, 436 Column Oven Heater, 230V, 436 Column Oven Heater, 120V, 456 Column Oven Heater, 230V, 456 CP745200 Main board for 436-GC and 456-GC LN2 Valve LCO2 Valve, DEFAULT Pressure relief Valve for LN2 option (only when present!) CP741079 Bulkhead assy NG-EFC 436-GC BR746195 Carriergas inlet manifold 456-GC Carriergas+ Make up Tee connection FERRULE, 1/16"ID VITON; PK/10

43 How to Order - Spares P/N DESCRIPTION CP91833104
SPI injector assy, 120V CP SPI injector assy, 230V 550023F Electronic Flowcontrol, type 23 Injector body EFC Injector Nut (=Septum nut) Septum Support Heater/Pt-100 Probe harness, 120V *) Heater/Pt-100 Probe harness, 230V *) Detector Body, FID Detector Base Igniter cable Signal cable 550011F Electronic Detector Flowcontrol, type 11 (DEFC-11) FID, PWA 4x6-GC (= Electrometer) CP745207 Detector interface board; 436/456-GC Pt-100 temperature probe Heater cartridge, 120V, 125W Heater cartridge, 230V, 125W Igniter Probe Signal Probe Collector Tube Insulator Screw, Micro seal, 8-32 Tower seals, 25/pk Flame Tip Assy, 0.02" ID, DEFAULT CR211200 Graphite ferrule 1/8" , 10/PK

44 Demonstrated Analysis

45 Results D2887a – N-Parafins Calibration
Peak ID Compound Name 1 n-C5 2 n-C6 3 n-C7 4 n-C8 5 n-C9 6 n-C10 7 n-C11 8 n-C12 9 n-C16 10 n-C18 11 n-C20 12 n-C22 13 n-C24 14 n-C28 15 n-C32 16 n-C36 17 n-C40 18 n-C44

46 Results D2887a – Reference GasOil
Peak ID Compound Name 1 n-C5 2 n-C6 3 n-C7 4 n-C8 5 n-C9 6 n-C10 7 n-C11 8 n-C12 9 n-C16 10 n-C18 11 n-C20 12 n-C22 13 n-C24 14 n-C28 15 n-C32 16 n-C36 17 n-C40 18 n-C44

47 Results D2887a – Blank Peak ID Compound Name 1 n-C5 2 n-C6 3 n-C7 4
9 n-C16 10 n-C18 11 n-C20 12 n-C22 13 n-C24 14 n-C28 15 n-C32 16 n-C36 17 n-C40 18 n-C44

48 Results D2887a – Sample Peak ID Compound Name 1 n-C5 2 n-C6 3 n-C7 4
9 n-C16 10 n-C18 11 n-C20 12 n-C22 13 n-C24 14 n-C28 15 n-C32 16 n-C36 17 n-C40 18 n-C44

49 Thank you

50 Pre-Install Requirements
Carrier gas Helium , > %, with Moisture and Oxygen filter Supply pressure 80 psi (5.3 bar) Detector gases FID Hydrogen (>99.999%, 40 psi / 2.6 bar) Air (moisture & oil free, 60 psi / 4 bar) Nitrogen (>99.999%, 80 psi / 5.3 bar). Charcoal filters for Helium, Hydrogen and Air are required.

51 Pre-Install requirements
Connections 1/8” Swagelok. (Cryogenic) Cooling Liquid 850–1000 psi (58.5 – 68.8 bar), with eductor tube. Compressed Air for Inlet 60 psi (4 bar). Gas Filters P/N Description #Required BRB1024B8 Filter kit for TCD incl. 2 filters (1/8” tube) 1 BRB1040S8 Filter kit for FID incl. 4 filters (1/8” tube) BRB0010B8 Connecting unit for 1 filter (1/8” tube) BRF0103 Hydrocarbon filter