Fig. 3. Modified Look–Locker inversion recovery (MOLLI) with 5(3)3 protocol. MOLLI method features several modifications intended to improve accuracy and.

Similar presentations


Presentation on theme: "Fig. 3. Modified Look–Locker inversion recovery (MOLLI) with 5(3)3 protocol. MOLLI method features several modifications intended to improve accuracy and."— Presentation transcript:

1 Fig. 3. Modified Look–Locker inversion recovery (MOLLI) with 5(3)3 protocol. MOLLI method features several modifications intended to improve accuracy and precision. This protocol employs two inversions and acquires three or five images after first and second inversions, with three RR intervals for T1 recovery. Images are sorted according to inversion time to perform pixel-wise three-parameter fitting. ECG = electrocardiogram, RR = the time interval between two consecutive R waves in the electrocardiogram, SSFP = steady-state free precession Fig. 3. Modified Look–Locker inversion recovery (MOLLI) with 5(3)3 protocol. MOLLI method features several modifications intended to improve accuracy and precision. This protocol employs two inversions and acquires three or five images after first and second inversions, with three RR intervals for T1 recovery. Images are sorted according to inversion time to perform . . . Korean J Radiol Jan-Feb;18(1):