Presentation is loading. Please wait.

Presentation is loading. Please wait.

Damiana Gentili Qualified Person/QU Director 09/05/2019

Similar presentations


Presentation on theme: "Damiana Gentili Qualified Person/QU Director 09/05/2019"— Presentation transcript:

1 Damiana Gentili Qualified Person/QU Director 09/05/2019
L’importanza della scelta del metodo analitico nella cleaning validation Damiana Gentili Qualified Person/QU Director 09/05/2019

2 Agenda Specific vs non-specific Visual inspection
Regulatory framework Specific vs non-specific Visual inspection Selection criteria for analytical methods & case history Analytical Validation overview Opportunities

3 FDA outcomes on analytical methods for CLEANING VALIDATION
Failure to adequately validate written procedures for the cleaning and maintenance of equipment

4 Italy

5 Analytical tecniques overview
Product Active ingredient Specific HPLC Mass Spectrometry Gas Chromatography Non-specific Total Organic Carbon UV-VIS Conductivity/pH Excipients

6 Analytical tecniques overview
Detergent Specific HPLC Mass Spectrometry Gas Chromtography Non-specific Total Organic Carbon UV-VIS Conductivity/pH Microbial Live microbes Bioburden Microbe breakdown Endotoxin

7 Detection of residues: from 1992 up to date… Specific vs. Non-specific

8 Specific vs. Non-specific Testing
Variety of residues Only one specific analyte Worst case results Expected by Regulatory Authority Highly sensitive Highly sensitive (sometimes) Less costly Costly Little sample preparation Sample preparation required Easy to manage Required technical skills Monitoring purposes Validation purposes

9 Characteristics Visible residue
This testing is designed to make observations directly from the clean equipment for quick turnaround Visual inspection using a visible residue limit (VRL) can monitor equipment cleaning efficiency The VRL is the level below which the residue of interest is not visible to the equipment Characteristics A VRL inspection is non-specific Any visible residue is considered unacceptable Several of its applications and associated risks have been demonstrate Since this is a limit test, the detection limit is the primary validation parameter. Precision should be also determined by using multiple observers in order to minimise subjectivity of the observers Subjectivity and training of the inspectors is the primary concern with this process Results from validation activity can be used to establish the difference between the Analytical Residue Limit and the Validation Residue Limit?

10 Analytical Methods - main criteria selection
The key factors in selecting a method are: Number of components to be analyzed Sampling technique Sample preparation processes Frequency and number of samples Data handling … besides sensitivity and selectivity

11 Case history : UPLC-MS Selection criteria:
High level of sensitivity requested due to a low LOD/ LOQ requested. LOQ in HPLC too high! High level of specificity requested due to 3 different contaminants to be detected Evaluation of sample preparation: Organic vs inorganic/soluble in water or other solvents Avoid sample manipulation (i.e. sample was in Pyridine/DMSO which would affect HPLC sensitivity) Swab recovery verification Is Analyte stable in the cleaning environment? Or does it degrade due to temperature during cleaning? Speed of analysis (throughput) as requirement  5 minutes analysis time Available instrumentation  yes, UPLC-MS (more than one) Operational cost: Time of QC development (pre-linearity, LOD and LOQ verification, swab recovery at 3 concentrations)  ab. 2 weeks

12 Final output Chromatographic conditions
The chromatographic procedure may be carried out using: - column: ACQUITY UPLC BEH C18, 100 × 2.1 mm, 1.7 μm or equivalent; - column temperature: 30°C; - Mobile phases: Mobile phase A: 0.1% formic acid in water (for example transfer 1.0 ml of formic acid 98% into 1000 ml water); Mobile phase B: 0.1% formic acid in acetonitrile (for example transfer 1.0 ml of formic acid 98% into 1000 ml water); - Detection: MS (settings are optimized for Waters Acquity QDa Detector), Single Ion Recording (SIR) Ionization Mode: ESI +; Probe Temp. 600 °C; Capillary 0.8 kV; Cone: 15 V; Acquisition Type: Selected Ion Recording (SIR); MASS: amu M984 MASS: amu N5.1 MASS: amu N5 - Injection volume: 1 µl; - Flow rate: 0.5 ml/min; - Run time: 5 min

13 Test method validation for CV samples
Testing cleaning validation samples requires a validated method The extent of validation is dependent upon the type of method employed the capabilities of the method the scientific and regulatory needs of the resulting data the anticipated outcome of the testing

14 Validation of Analytical Methods
It is essential to validate the analytical methods chosen for cleaning validation and routine monitoring in the ranges encountered during cleaning The minimum requirement for method validation must include the following tests: LOQ. In practice it may not be necessary to determine the LOQ but to show that it sufficiently low for the current purposes when sensitive techniques are being used LOD Specificity for chromatographic or UV methods Accuracy Repeatability Recovery from spiked surfaces Stability of the analyte in the cleaning solvent

15 Where we are as API manufacturer
We use specific analytical method rather than non-specific We use solvents as cleaning agents and very low use of detergents We use high level of sophisticated equipment NEW OPPORTUNITIES TO EXPLORE Use non-specific methods/easiest analytical tecniques for monitoring purposes Use detergents to reduce safety issues in handling solvents as cleaning agents Correlate monitoring results with validation data Increase the level of control of cleaning

16 Thank you very much gentili@procos.it
Tempio del padiglione d'oro Thank you very much


Download ppt "Damiana Gentili Qualified Person/QU Director 09/05/2019"

Similar presentations


Ads by Google