Wien filter 2 06/05/2019 Gabriel Palacios
2
Details Solidworks model CST simulations |E| Ey Ez
Comparing 20 kV vs 40 kV feedthroughs
3
Solidworks model
4
Solidworks model This 40 kV feedthrough prevents us from having to change the pipe to the 1.5” OD that interferes with the coils. The Kovar recess measurements where “eyeballed”. The HV spring guide connector requires mods. 0.06” gap Between 1” OD and 0.75 OD insulator
5
Solidworks model The tip rounded could be removed to improve contact with feedthrough Kovar recess.
6
Solidworks model Modeled in Solidworks 40 KV Single Pin CF
7
Solidworks model Modeled in Solidworks
Based on Electrical feedthrough CAD
8
CST simulations: electric field norm |E|
10 [MV/m] scale max
9
CST simulations: electric field norm |E|
~5 [MV/m] ~5.4 [MV/m] ~6.8 [MV/m] 10 [MV/m] scale max ~4.1 [MV/m]
10
CST simulations: electric field norm |E|
10 [MV/m] scale max 2.6 [MV/m] at cusp 2.5 [MV/m] at cusp
11
CST simulations: electric field norm |E|
5 [MV/m] scale max
12
CST simulations: electric field norm |E|
5 [MV/m] scale max
13
CST simulations: electric field norm |E|
5 [MV/m] scale max
14
CST simulations: electric field vertical Ey
3 [MV/m] scale max
15
CST simulations: electric field vertical Ey
~2.6 [MV/m] red region ~0.8 [MV/m] ~2.3 [MV/m] blue region 3 [MV/m] scale max
16
CST simulations: electric field vertical Ey
3 [MV/m] scale max ~2.7 [MV/m] at cusp 1.4 [MV/m] at red region -2 [MV/m] at blue region
17
CST simulations: electric field longitudinal Ez
5 [MV/m] scale max
18
CST simulations: electric field longitudinal Ez
~-3.1 [MV/m] 5 [MV/m] scale max ~-5.5 [MV/m]
19
CST simulations: electric field longitudinal Ez
5 [MV/m] scale max ~-1.7 [MV/m] at cusp ~-2.5 [MV/m] at cusp
20
CST simulations: electric field norm |E|
~5 [MV/m] ~10 [MV/m] red region ~5.4 [MV/m] ~5 [MV/m] ~6.8 [MV/m] ~4.3 [MV/m] ~4.1 [MV/m] ~5.2 [MV/m]